Effect of high frequency oscillatory high flow nasal cannula on carbon dioxide clearance in a premature infant lung
Emidio M Sivieri1,2, Eric Eichenwald2,3, Salma M Bakri1
1CHOP Newborn Care at Pennsylvania Hospital, Philadelphia, Pennsylvania.
Objective:
This study compared CO2 clearance in a premature infant lung model connected to a high flow nasal cannula (HFNC) system supplied with oscillatory versus non-oscillatory flow.
Design/Methods:
The lung model was set to compliance 1.0 mL/cmH2 O, RR 60 breaths/min, and 6 mL tidal volume. A 100% CO2 was injected at a constant 15 mL/min. To create oscillation, HFNC flow was interrupted at rates of 4-6-8 and 10 Hz. equilibrated end-tidal CO2 (ETCO2 ) was recorded with and without oscillation at set flows of 2-8 L/min and repeated for each oscillation frequency.
Results:
Overall ETCO2 decreased significantly (P < 0.001) during both non-oscillatory and oscillatory HFNC as set flow increased from 2 to 8 L/min by 26.3% and 60.8%, respectively. Oscillatory ETCO2 levels decreased linearly compared to non-oscillatory HFNC with negligible difference at 2 L/min and a 48.4% difference at 8 L/min (P < 0.001). There were no differences in ETCO2 levels between oscillation frequencies at any flow except at 6 Hz for which ETCO2 was significantly lower (P < 0.01) than at 4, 8, and 10 Hz for 5-8 L/min HFNC flows. Amplitude of volume oscillations increased with increasing flow from 0.5 mL at 2 L/min to 4.0 mL at 8 L/min (P < 0.001), and decreased with increasing oscillation frequency.
Conclusion:
Oscillatory HFNC as compared to non-oscillatory was associated with significantly improved CO2 clearance in this premature infant lung model. This simple modification of the HFNC system may prove to be a useful enhancement to this mode of non-invasive respiratory support for preterm infants at high risk for respiratory failure.
More Related Videos
11:26Assessing Cerebral Autoregulation via Oscillatory Lower Body Negative Pressure and Projection Pursuit Regression
Published on: December 10, 2014
11:13Collection, Expansion, and Differentiation of Primary Human Nasal Epithelial Cell Models for Quantification of Cilia Beat Frequency
Published on: November 10, 2021
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Carbon-dioxide Fixation
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Carbon Dioxide Transport in the Blood
Forms of CO2 Transport
1. Dissolved in plasma: A small percentage (7-10%) of CO2 is transported and dissolved directly in the plasma.
2. Carbaminohemoglobin: Just over 20% of CO2 is chemically bound to...
Clearance Models: Physiological Models
The organ's clearance rate depends on the blood flow to the organ and the extraction ratio (E). The extraction ratio describes the organ's...
Clearance Models: Compartment Models