Fat volume measurements as a predictor of image noise in coronary computed tomography angiography
Rami M Abazid1, Osama A Smettei1, Ahmad Almeman2
1Department of Cardiology, Cardiac Imaging, Prince Sultan Cardiac Center-Qassim PSCCQ, King Fahad Specialist Hospital, Buraydah, Saudi Arabia.
Introduction:
Image noise can negatively affect the overall quality of coronary computed tomography angiography (CCTA).
Objectives:
The purpose of this study was to evaluate the relationship between image noise and fat volumes in the chest wall. We also aimed to compare these with other patient-specific predictors of image noise, such as body weight (BW) and body mass index (BMI).
Methods:
We undertook a cross-sectional, single-center study. A tube voltage of 100 kV was used for patients with BW <85 kg and 120 kV for BW ≥85 kg. The image noise in the aortic root, single-slice fat volume (SFV) at the level of the left main coronary artery and the total fat volume of the chest (TFV) were analyzed.
Results:
A total of 132 consecutive patients were enrolled (mean age ± standard deviation, 51 ± 11 years; 64% male). The mean image noise was 30.5 ± 11 Hounsfield units (HU). We found that patients with image noise >30 HU had significantly higher SFV (75 ± 33 vs. 51 ± 24, p < 0.0001) and TFV (2206 ± 927 vs. 1815 ± 737, p < 0.01) compared with patients having noise ≤30 HU, whereas BW and BMI showed no significant difference (78 ± 13 vs. 81 ± 14, p < 0.34) and (28.7 ± 4.7 vs. 26.8 ± 3.8, p < 0.19), respectively. Linear regression analysis showed that image noise has better correlation with SFV (R = 0.399; p < 0.0001); and TFV (R = 0, p < 0.009) than BMI (R = 0.154, p < 0.039) and BW (R = -0.102, p = 0.12).
Conclusions:
Fat volume measurements of the chest wall can predict CCTA image noise better than other patient-specific predictors, such as BW and BMI.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Coronary Artery Disease IV: Preventive Measures
Coronary Circulation
Coronary circulation begins at the base of the aorta, where two main arteries arise—the left and right coronary arteries. These arteries encircle the heart in the coronary sulcus and supply the...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Constant Volume Calorimetry