[Study on the Imaging Quality Evaluation of Aspheric Intraocular Lenses]
Qin Feng1, Hong Lu2, Yongjie Luo1
1Zhejiang Institute of Medical Device Testing, Hangzhou, 310019.
Abstract:
The article is to study the objective evaluation method of the imaging quality of aspheric intraocular lens with different design concepts by establishing the model eye which is close to the actual corneal spherical aberration characteristics. Combined with the research data of corneal spherical aberration, the simulated cornea corresponding to Zernike c(4,0) coefficient 0.00 μm, 0.10 μm, 0.20 μm and 0.28 μm was established respectively. The optical quality of the measured simulated corneal lenses basically reached the theoretical design effect, and the deviation between the MTF value and the theoretical design value was equivalent to the MTF test repeatability. The establishment of model eye for artificial lens provides a research tool for the evaluation of aspheric intraocular lens imaging quality.