A Preliminary Proposal for Quality Control Assessment and Harmonization of Leukocytes Morphology-structural
Michela Seghezzi1, Sabrina Buoro1, Giulia Previtali1
1Clinical Chemistry Laboratory, Hospital Papa Giovanni XXIII, Bergamo, Italy.
Background:
The cell population data (CPD) measured by Sysmex XN-9000 can be used for screening many hematological and non-hematological disorders. Since little information is available on harmonization of CPD among different instrumentation and clinical laboratories, this study aimed at assessing the current degree of CPD harmonization between separate Sysmex XN modules allocated to the same laboratory.
Methods:
A total number of 78291 data were used for verification of within-run imprecision, analyzers harmonization, reference ranges and assessment of blood sample stability of CPD parameters, including results of daily quality control testing and those generated in samples collected from blood donors and healthy volunteers.
Results:
Within-run imprecision of CPD parameters ranged between 0.4 and 14.1%. Good agreement was found among five different XN-modules, especially when values were adjusted after calculation of instrument-specific alignment factors. The bias of all parameters remained always lower than the reference change values in samples stored for up to 8 hours, regardless of storage temperature.
Conclusions:
The imprecision of CPD parameters was acceptable, except for those reflecting the dispersion of cellular clusters. Due to the lack of reference control materials, we showed that the use of data generated on a large number of normal routine samples (i.e., a Moving Average population) may be a reliable approach for testing analyzers harmonization. Nevertheless, availability of both calibration and quality control materials for these parameters is highly advisable in the future. We finally showed that whole blood samples may be stable for up to 2-4 hours for most CPD parameters.
More Related Videos
08:56Clinical Assessment of Spatiotemporal Gait Parameters in Patients and Older Adults
Published on: November 7, 2014
08:43Author Spotlight: Standardizing Spheroid Formation Methods for Metabolic and Oxygenation Analysis Using Fluorescence Lifetime Imaging Microscopy
Published on: August 9, 2024
Related Concept Videos
Distributions to Estimate Population Parameter
Wave Parameters
Model Approaches for Pharmacokinetic Data: Distributed Parameter Models
The distributed parameter models are specifically designed to account for variations and differences in some drug classes. This model is particularly useful for assessing regional concentrations of anticancer or...
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Noncompartmental Analysis: Miscellaneous Pharmacokinetic Parameters
One key aspect of the noncompartmental approach is determining a drug's total clearance. This can be done by dividing the drug dose by the area under the concentration-time curve from zero to infinity. The area under the concentration-time curve represents the drug's...
Dosage Regimens: Partial Pharmacokinetic Parameters