Prediction of Red Blood Cell Antibody Significance Using the Monocyte-Macrophage Assay
Published on: February 7, 2025
Comparison of two column agglutination tests for red blood cell antibody testing
Jonas Sawierucha1, Marion Posset1, Viola Hähnel1
1Institute for Clinical Chemistry and Laboratory Medicine, Transfusion Medicine, University Hospital Regensburg, Regensburg, Germany.
Background:
Several sensitive methods are available for red blood cell (RBC) antibody screening. Among these, gel and glass card systems have demonstrated comparably good performance in retrospective studies and are widely used in routine patient diagnostics, but their performance in prospective studies has not been sufficiently characterised.
Patients And Methods:
Gel card (Bio-Rad DiaMed) and glass bead-based (Ortho Clinical Diagnostics) column agglutination technologies were used to screen for antibodies prospectively (group A) and for antibody identification in stored and fresh samples known to contain RBC antibodies retrospectively (group B). Untreated reagent RBCs and either papain-treated (Bio-Rad) or ficin-treated panel C cells (Ortho) were used for antibody identification.
Results:
RBC-reactive antibodies were detected in 22 of 1000 group A samples, three of which tested positive only by gel card agglutination, and four only by glass bead agglutination (including one false positive each). Group B comprised 202 sera with known antibodies: 33 of these samples contained 36 antibodies detected only by gel card agglutination, whereas 9 samples contained antibodies detectable only by glass bead-based agglutination. Discrepancies mostly involved weak antibodies reactive by enzyme only. Two sera contained antibody mixtures that neither system detected completely. Of note, in antibody differentiation batches one and two, anti-Lua was reactive in 7 of 7 and 1 of 8 samples, respectively.
Conclusion:
Both column agglutination tests for red cell antibodies had equal sensitivity and specificity with unstored samples. In stored samples, weak and enzyme-only antibodies were more frequently detected with the gel card system.
Related Concept Videos
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Blood Transfusion and Agglutination
History
The history of blood transfusion dates back to the 17th century, when early attempts were made in animals. In 1818 James Blundell, a British doctor, performed the first successful human blood transfusion. Later in 1900, Karl...
Testes: Histology
The spermatogenic cells, responsible for producing sperm, are...
Complementation Tests
Organisms heterozygous for different mutations are crossed pairwise in all combinations. If present on different genes, the mutations can complement each other by providing the missing...
Significance Testing: Overview
Testes: Gross Anatomy
Each testis is surrounded by the tunica albuginea, a dense connective tissue layer that provides structural support and protection. This layer is covered by an outer serous membrane called the tunica vaginalis, which helps reduce friction...