Impact of body part thickness on AP pelvis radiographic image quality and effective dose
K Alzyoud1, P Hogg2, B Snaith3
1School of Health Sciences, University of Salford, Salford M6 6PU, United Kingdom; Faculty of Allied Health Sciences, Hashemite University, Zarqa, Jordan.
Introduction:
Within medical imaging variations in patient size can generate challenges, especially when selecting appropriate acquisition parameters. This experiment sought to evaluate the impact of increasing body part thickness on image quality (IQ) and effective dose (E) and identify optimum exposure parameters.
Methods:
An anthropomorphic pelvis phantom was imaged with additional layers (1-15 cm) of animal fat as a proxy for increasing body thickness. Acquisitions used the automatic exposure control (AEC), 100 cm source to image distance (SID) and a range of tube potentials (70-110 kVp). IQ was evaluated physically and perceptually. E was estimated using PCXMC software.
Results:
For all tube potentials, signal to noise ratio (SNR) and contrast to noise ratio (CNR) deceased as body part thickness increased. 70 kVp produced the highest SNR (46.6-22.6); CNR (42.8-17.6). Visual grading showed that the highest IQ scores were achieved using 70 and 75 kVp. As thickness increases, E increased exponentially (r = 0.96; p < 0.001). Correlations were found between visual and physical IQ (SNR r = 0.97, p < 0.001; CNR r = 0.98, p < 0.001).
Conclusion:
To achieve an optimal IQ across the range of thicknesses, lower kVp settings were most effective. This is at variance with professional practice as there is a tendency for radiographers to increase kVp as thickness increases. Dose reductions were experienced at higher kVp settings and are a valid method for optimisation when imaging larger patients.
More Related Videos
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Drug Control Governance: Regulatory Bodies and Their Impact
Veins of the Abdomen and Pelvis
The inferior vena cava is fed by numerous smaller veins. The lumbar veins, for instance, drain the posterior abdominal wall, emptying both directly into the inferior vena cava and into the...
Biological Effects of Radiation
Impact of Groups on Groups
Framing Effects