Radiation dose reduction using two orthogonal topograms associated with automatic tube voltage selection for lung CT
Xiaofei Liu1, Mingming Zhang2, Liang Zhang3
1Department of Radiology, The First Hospital of China Medical University, No. 155 Nanjingbei Street, Shenyang, 110001, Liaoning, People's Republic of China.
Objective:
To evaluate the effectiveness of two orthogonal topograms on radiation dose and image quality (IQ) associated with topogram-based automatic tube voltage selection (ATVS) for lung CT scanning.
Methods:
Thirty-seven patients were enrolled in this study. At baseline, only an anteroposterior topogram was obtained and at follow-up, both anteroposterior and lateral topograms were performed. ATVS was turned on during all scans. Objective and subjective IQ evaluations were performed and compared; tube voltage and radiation dose of each scan were noted and analyzed.
Results:
A significant difference was observed regarding the objective parameters between baseline and follow-up only in image noise and signal-noise ratio (SNR) in the upper one-third of the image (image noise: 7.49 ± 1.08 vs. 9.10 ± 1.13, p < 0.001; SNR: 4.08 ± 0.87 vs. 3.37 ± 0.63, p < 0.001). No differences were found between baseline and follow-up in the subjective assessment of IQ. The radiation dose was significantly lower at follow-up than that at baseline (2.73 ± 0.83 mSv vs. 3.55 ± 1.24 mSv, respectively; p < 0.001).
Conclusions:
Using two orthogonal topograms associated with ATVS could significantly reduce the total radiation dose for lung CT scanning, while subjective IQ was maintained.