Dosimetric study of CO-60 source step size in uterine cervix intracavitary HDR brachytherapy
Arvind Kumar Shukla1, Bhupendra Singh Rana2, Narinder Paul Singh3
1Department of Applied Sciences, IK Gujral Punjab Technical University (PTU), Jalandhar, India; Department of Radiotherapy, RNT Medical College, Udaipur, India.
Purpose:
The present work reports effects of source step sizes on dose distribution in patients treated with cobalt-60 (Co-60) high-dose-rate afterloading brachytherapy in carcinoma cervix (Ca-cx).
Methods And Materials:
The retrospective study is based on data of 15 patients of Ca-cx treated with Co-60 high-dose-rate intracavitary brachytherapy with dose of 21 Gy in three fractions with source step size of 2.5 mm after external beam radiotherapy of 46 Gy. The effect of source step size on overall treatment procedure was evaluated from prescribed dose volume, dose to organ at risks, and treatment time for source step sizes of 1 mm, 2.5 mm, 5 mm, and 10 mm for each patient.
Results:
The mean dose to bladder point for 1 mm, 2.5 mm, 5 mm, and 10 mm source step sizes was found to be 3.37 Gy (SD: 1.36), 3.44 Gy (SD: 1.38), 3.54 Gy (SD: 1.41), and 3.74 Gy (SD: 1.46), respectively. Similarly, the mean dose received by rectum point for these source step sizes were 2.86 Gy (SD: 0.64), 3.02 Gy (SD: 0.67), 3.25 Gy (SD: 0.71), and 3.63 Gy (SD: 0.73), respectively. The treatment time and prescribed dose coverage volume were both found to be gradually increasing with increase in step size.
Conclusions:
Our results on Ca-cx brachytherapy using Co-60 source indicate that the prescribed dose volume gradually increases from smaller source step to larger source step size. This results in increase of dose to the bladder and rectum and may lead to increase in toxicity and reduces quality of life. The study recommends that step size more than 5 mm should not be used for uterine cervix intracavitary application using Co-60 source.
Related Concept Videos
Uterus and Cervix
The uterus is securely anchored within the pelvic cavity by paired broad ligaments on either side. It is further stabilized by three pairs...
Uterine Tubes
Genome Size and the Evolution of New Genes
Cell Size
Surface Area
Cells can take in nutrients and water via diffusion through the plasma membrane itself or through specific channels in the membrane. The area of the membrane surrounding...
Rate-Determining Steps
In a multistep reaction mechanism, one of the elementary steps progresses significantly slower than the others. This slowest step is called the rate-limiting step (or rate-determining step). A reaction cannot proceed faster than its slowest step, and hence, the rate-determining step limits the overall reaction rate.
The concept of rate-determining step can be understood from the analogy of a 4-lane freeway with a short-stretch of traffic-bottleneck caused due to...
Sinusoidal Sources
In homes, the power supplies use sinusoidal sources to provide electricity. These sources generate a voltage that varies sinusoidally...