Optimizing a Portable ERG Device for Glaucoma Clinic: The Effect of Interstimulus Frequency on the Photopic Negative
Flora Hui1, Jessica Tang1,2, Xavier Hadoux1
1Centre for Eye Research Australia, Royal Victorian Eye and Ear Hospital, Melbourne, VIC, Australia.
Purpose:
The purpose of this study was to investigate the effect of interstimulus frequency on the photopic negative response (PhNR) in the clinical electroretinogram (ERG) in glaucoma and healthy eyes.
Methods:
Participants with open angle glaucoma (n = 15) and age-matched controls (n = 20) were recruited. Photopic ERGs were recorded in one eye using five frequencies (1-5 Hz) delivered in random order. ERGs were analyzed for changes to amplitude and timing between groups and interstimulus frequency. Coefficient of variation (CoV) was used to examine variability within recordings for each frequency.
Results:
While the a-wave and b-wave showed minimal alteration, the PhNR was highly sensitive to changes in interstimulus frequency. The PhNR signal was largest at 1 Hz and steadily diminished with higher frequencies in both control and glaucoma groups. Significant differences in PhNR amplitude were found between controls and glaucoma groups at 2 and 3 Hz. While 1 Hz delivered the largest PhNR, it also showed a significantly greater CoV compared to other frequencies.
Conclusions:
An interstimulus frequency of 2 Hz was optimal for recording the PhNR, creating a good balance between testing time and signal quality. A higher frequency could be used to further shorten clinical testing times; however, this may compromise its clinical utility by dampening the PhNR.
Translational Relevance:
Here we show the importance of considering flash interstimulus frequency when designing ERG protocols for recording the PhNR as while higher frequencies can shorten test times, they also have considerable effects on the PhNR.
Related Concept Videos
Frequency Response of BJT
Low-Frequency Response: At low frequencies, the behavior of the BJT is determined by its DC bias point, which is set by the...
Frequency Response of a Circuit
The transfer function is pivotal in characterizing how these circuits react to various frequencies, facilitating a profound understanding of their behavior. An essential parameter is the time constant, signifying the...
Frequency Response of Op Amp Circuits
Frequency Response and Gain:
The gain of the op-amp, A(ω), is not a constant but a function of the input signal frequency. An op-amp can maintain a constant gain at low frequencies, known...
Glaucoma: Overview
Frequency-dependent Selection
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...