Evaluation of the B-Smart manometer and the CompuFlo computerized injection pump technology for accurate needle-tip
Robyn S Weisman1, Nirav P Bhavsar2, Kathleen A Schuster2
1Department of Anesthesiology, University of Miami - Miller School of Medicine, Miami, Florida, USA rweisman1@med.miami.edu.
Background And Objectives:
The exact mechanism of peripheral nerve blocks causing/leading to nerve injury remains controversial. Evidence from animal experiments suggests that intrafascicular injection resulting in high injection pressure has the potential to rupture nerve fascicles and may consequently cause permanent nerve injury and neurological deficits. The B-Smart (BS) in-line manometer and the CompuFlo (CF) computerized injection pump technology are two modalities used for monitoring pressure during regional anesthesia. This study sought to explore the accuracy of these two technologies in measuring needle-tip pressures in a simulated environment.
Methods:
In seven simulated needle-syringe combinations, the BS and the CF devices were connected in series through a closed system and attached to a digital manometer at the tip of various needles. The pressures were evaluated in three trials per needle-syringe combination. Sensitivity, specificity, positive predictive value, negative predictive value, and accuracy (F1 Score) were determined for each needle type and overall.
Results:
For pressures ≥15 psi and ≥20 psi, respectively, the CF device demonstrated a sensitivity of 100%, 100%; specificity of 96%, 98%; positive predictive value 93%, 93%; and negative predictive value of 100%, 100%. The BS device demonstrated a sensitivity of 60%, 100%; specificity of 99%, 95%; positive predictive value of 96%, 85%; and negative predictive value of 85%, 100%. Accuracy, as measured by the F1 Score, for detecting a pressure of ≥15 psi was 0.96 for the CF and 0.74 for the BS.
Conclusions:
Future research is needed to explore in-vivo performance and evaluate whether either of these devices can impact on clinical outcomes.
Related Concept Videos
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer
Peripheral Nervous System: Ganglia and Nerves
Nerves
The nerve is a bundle of axons that serves as the communication highway in the PNS. Each nerve is ensheathed in a protective layer of connective tissue called the epineurium. This outermost layer safeguards the nerve and supports the...
Neuromuscular Junction And Blockade
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Measurement of Blood Pressure
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...