Seroma Visualization and Implant Accuracy in Permanent Breast Seed Implant Brachytherapy
Daniel Morton1, Michelle Hilts2, Deidre Batchelar3
1Department of Medical Physics, BC Cancer - the Southern Interior, Kelowna, British Columbia, Canada; Department of Physics and Astronomy, University of Victoria, Victoria, British Columbia, Canada.
Purpose:
This study aimed to evaluate the relationship between seroma visualization and seed placement accuracy in permanent breast seed implant brachytherapy (PBSI).
Methods And Materials:
At the time of planning computed tomography (CT), 10 patients receiving PBSI were imaged with spatially co-registered 3-dimensional ultrasound (US). Seromas were independently contoured by 3 radiation oncologists on CT and US scans. Intra- and interuser conformity indices (CIs) were used as a surrogate for seroma visualization. Intermodality visualization differences were assessed by defining consensus contours, clinical target volume (CTV)CT and CTVUS, and evaluating the CI and the centroid position and volume differences. Seed placement accuracy was represented by the differences between the planned and implanted seed positions (displacements). Correlations among total, systematic, and random seed displacements and seroma visualization metrics were assessed.
Results:
The median (range) intra-user CI of CT seroma contouring was 0.60 (0.46-0.72), and the median interuser CIs were 0.46 (0.38-0.58) and 0.50 (0.29-0.67) on CT and US, respectively. The CTVUS was a mean 68% ± 12% smaller than CTVCT and differed in centroid position by 8 ± 3 mm. Seeds were placed, on average, 10 ± 5 mm from their planned positions, and intrapatient systematic displacements were observed. The mean seed displacements for the implants were shown to correlate with interuser CI on CT (r = .74; P = .01) and volume differences between CTVCT and CTVUS (r = .65; P = .04), but not with intrauser CI, intermodality CI or centroid differences. Systematic displacements were correlated with interuser CT CI (r = .67; P = .03) and intermodality volume difference (r = .64; P = .04), but random seed displacements were independent of all evaluated metrics.
Conclusions:
Consistency in seroma delineation in treatment planning and differences between seroma visualized on CT and US scans are associated with seed placement accuracy in PBSI. Efforts to enhance seroma visualization in treatment planning and implant guidance may have a positive impact on treatment quality and should be pursued to facilitate the widespread implementation of this technique.
Related Concept Videos
Improving Translational Accuracy
Improving Translational Accuracy
Uncertainty in Measurement: Accuracy and Precision
Seed Structure and Early Development of the Sporophyte
Accuracy and Precision
Accuracy, limits, and approximation
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...