On the application of balanced steady-state free precession to MR microscopy
Sébastien Bär1,2,3, Thomas Oerther4, Matthias Weigel5,6
1Department of Radiology, Medical Physics, Faculty of Medicine, Medical Center - University of Freiburg, Killianstrasse 5a, 79106, Freiburg, Germany. sebastien.baer@uniklinik-freiburg.de.
Objective:
The applicability of the balanced steady-state free precession (bSSFP) sequence to the field of MR microscopy was investigated, since the potentially high SNR makes bSSFP attractive. However, particularly at ultra-high magnetic fields, a number of constraints emerge: the frequency sensitivity of the bSSFP signal, the duty cycle of the imaging gradients, and the intrinsic diffusion attenuation of the steady state due to the imaging gradients.
Materials And Methods:
Optimization of the bSSFP sequence was performed on three imaging systems (7 T and 9.4 T) suited for MR microscopy. Since biological samples are often imaged in the very proximity of materials from sample containers/holder or devices such as electrodes, several microscopy phantoms representing such circumstances were fabricated and examined with 3D bSSFP.
Results:
Artifact-free microscopic bSSFP images could be obtained with voxel sizes down to 16 µm × 16 µm × 78 µm and with an SNR gain of 25% over standard gradient echo images.
Conclusion:
With appropriate choice of phantom materials, optimization of the flip angle to the diffusion-attenuated steady state and protocols considering duty-cycle limitations, bSSFP can be a valuable tool in MR microscopy.
More Related Videos
Related Concept Videos
Gyroscope: Precession
Balancing Redox Equations
Steady State Concentration
Most drugs are administered in repeated doses at fixed intervals or through continuous...
Atomic Nuclei: Larmor Precession Frequency
Steady Flow of a Fluid Stream
During this process, the momentum of the fluid within the control volume remains constant over the time interval dt. By applying the...
Transient and Steady-state Response
These test signals are integral in designing control systems to exhibit two key performance aspects: transient response and steady-state...