Related Experiment Video
Updated: Sep 2, 2026

High-Pressure NMR Experiments for Detecting Protein Low-Lying Conformational States
Published on: June 29, 2021
Probe with four independent detectors for parallel detection in high-resolution solution-state NMR
Sebastiaan Van Dyck1, Coline Wiame1, Hadi Loutfi1
1Chimie Physique et Chimie du Vivant (CPCV, UMR 8228), Department of Chemistry, Ecole Normale Supérieure, PSL University, Sorbonne University, CNRS, Paris 75005, France.
Abstract:
A quadruple probe was designed for wide-bore spectrometers with four independent detectors, with sample volumes of ∼5 μl each. Four transmission-line detectors are equipped with radio frequency channels for 1H (or 19F) and 13C, which can be tuned and matched independently. Each detector is equipped with 6 on-chip shim-coils for local shimming. Linear pulsed field gradients can be applied independently to each sample along the z-axis (in the B0 direction). Slice selection allows one to reduce the effective volumes to ∼2.5 μl to improve the resolution and minimize crosstalk. Four sets of experiments can be acquired simultaneously on four different samples, for example, four inversion-recovery or four Carr-Purcell-Meiboom-Gill experiments.
More Related Videos
08:0915N CPMG Relaxation Dispersion for the Investigation of Protein Conformational Dynamics on the µs-ms Timescale
Published on: April 19, 2021
12:47Nuclear Magnetic Resonance Spectroscopy for the Identification of Multiple Phosphorylations of Intrinsically Disordered Proteins
Published on: December 27, 2016
Related Concept Videos
Mass Analyzers: Common Types
2D NMR: Overview of Homonuclear Correlation Techniques
COSY90 is the standard two-dimensional (2D) COSY experiment that...
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
Interpreting ¹H NMR Signal Splitting: The (n + 1) Rule
NMR Spectrometers: Resolution and Error Correction
NMR Spectrometers: Overview