Treatment of Facial Deformities using 3D Planning and Printing of Patient-Specific Implants
Published on: May 23, 2020
Patient-specific 3D printed plates improve stability of Le Fort 1 osteotomies in vitro
Kasper Stokbro1, Søren Wiatr Borg2, Morten Østergaard Andersen3
1Department of Oral and Maxillofacial Surgery, Odense University Hospital, Sdr. Boulevard 29, 5000, Odense, Denmark; Department of Clinical Institute, Faculty of Health, University of Southern Denmark, Winsløwparken 19, 5000, Odense, Denmark.
Purpose:
Selective laser melting used to manufacture patient-specific 3D-printed (PSP) plates is a delicate process, which may introduce weakened areas in the plates, with risk of fracture. This in vitro study's purpose was to test the ability of PSP plates to stabilize Le Fort I osteotomies compared with manually adapted stock plates. The study's objectives were to measure the force needed to compress the osteotomy and evaluate whether the PSP plates would break during compression.
Materials And Methods:
This controlled in vitro study evaluated the maxillary stability using the clinical data from 7 patients. The virtually planned maxillary reposition was 3D-printed in 2 copies, and the osteotomy gap was fixated by either PSP plates or stock plates. The models were compressed until the Le Fort I osteotomy gap was eliminated. The primary outcome was the force needed to compress the model. The primary predictor variable was a comparison between PSP and stock plates. Secondary outcome measurements were the slope of elastic modulus, yield point, and force needed for 2 mm compression. Statistical testing was performed by Wilcoxon signed-rank test with significance level at P ≤ 0.05.
Results:
The PSP plates performed better than stock plates in all outcome measurements. None of the plates broke during compression despite forces of more than 4000 N. The first point of failure in PSP plates was the first screw cranial to the osteotomy. In comparison, the first point of failure in stock plates was in the plates' bend at the osteotomy.
Conclusion:
In this in vitro setup, the Le Fort I osteotomies fixated with PSP plates were more stable than the osteotomies fixated with conventional stock plates. No adverse effects occurred during testing of PSP plates; thus, PSP plates seem to be a safe alternative to stock plates and may even be preferable.
More Related Videos
Related Concept Videos
Le Chatelier's Principle: Changing Concentration
Le Chatelier's Principle: Changing Temperature
To understand this phenomenon, consider the elementary reaction:
Le Chatelier's Principle: Changing Volume (Pressure)
RNA Stability
Nuclear Stability
To hold positively charged protons together...
Stability
The stability of an LTI system is determined by the roots of its characteristic equation, known as poles. A system is stable if it produces a bounded...