A Hand-Held Device to Apply Instrument-Assisted Soft Tissue Mobilization at Targeted Compression Forces and Stroke
John B Everingham1, Peter T Martin1, Trevor J Lujan2
1Department of Mechanical and Biomedical Engineering, Boise State University, 1910 University Drive, Boise, ID 83725-2085.
Abstract:
Instrument-assisted soft tissue mobilization (IASTM) is a manual therapy technique that is commonly used to treat dysfunctions in ligaments and other musculoskeletal tissues. The objective of this study was to develop a simple hand-held device that helps users accurately apply targeted compressive forces and stroke frequencies during IASTM treatments. This portable device uses a force sensor, tablet computer, and custom software to guide the application of user-specified loading parameters. To measure performance, the device was used to apply a combination of targeted forces and stroke frequencies to foam blocks and silicone pads. Three operators using the device applied targeted forces between 0.3 and 125 N with less than 10% error and applied targeted stroke frequencies between 0.25 and 1.0 Hz with less than 3% error. The mean error in applying targeted forces increased significantly at compressive forces less than 0.2 N and greater than 125 N. For experimental validation, the device was used to apply a series of IASTM treatments over three-weeks to rodents with a ligament injury, and the targeted compressive force and stroke frequency were repeatedly applied with an average error less than 5%. This validated device can be used to investigate the effect of IASTM loading parameters on tissue healing in animal and human studies, and therefore can support the optimization and adoption of IASTM protocols that improve patient outcomes.
More Related Videos
13:44Haptic/Graphic Rehabilitation: Integrating a Robot into a Virtual Environment Library and Applying it to Stroke Therapy
Published on: August 8, 2011
07:34Functional MRI in Conjunction with a Novel MRI-compatible Hand-induced Robotic Device to Evaluate Rehabilitation of Individuals Recovering from Hand Grip Deficits
Published on: November 23, 2019
Related Concept Videos
Uncertainty in Measurement: Reading Instruments
Frequency-dependent Selection
Intermolecular Forces in Solutions
When the strengths of the intermolecular forces of attraction between solute and solvent species in a solution are no different than those present in the separated components, the solution is formed with no accompanying energy change. Such a solution is called an ideal solution. A mixture of ideal gases (or gases such as helium and argon,...
Hand hygiene
Hand washing...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...