Clinical Assessment of Metal Artifact Reduction Methods in Dual-Energy CT Examinations of Instrumented Spines
Zaiyang Long1, David R DeLone1, Amy L Kotsenas1
11 Department of Radiology, Mayo Clinic, 200 1st St SW, Rochester, MN 55905.
Objective:
The purpose of this study was to evaluate the performance of three metal artifact reduction methods in dual-energy CT (DECT) examinations of instrumented spines.
Materials And Methods:
Twenty patients with instrumented spines who underwent spine DECT were retrospectively identified. All scans were obtained on a dual-source 128-MDCT scanner. In addition to the original DE mixed images, DECT images were reconstructed using an iterative metal artifact reconstruction algorithm (DE iMAR), virtual monochromatic imaging (VMI) algorithm (DE Mono+), and a combination of the two algorithms DE iMAR and DE Mono+, which we refer to here as "DE iMAR Mono+." The four image series were anonymized and randomized for a reader study. Four experienced neuroradiologists rated the images in terms of artifact scores of four anatomic regions and overall image quality scores in both bone and soft-tissue display window settings. In addition, a quantitative analysis was performed to assess the performance of the three metal artifact reduction methods.
Results:
There were statistically significant differences in the artifact scores and overall image quality scores among the four methods (both, p < 0.001). DE iMAR Mono+ showed the best artifact scores and quality scores (all, p < 0.001). The intraclass correlation coefficient for the overall image quality score was 0.779 using the bone display window and 0.892 using the soft-tissue display window (both, p < 0.001). In addition, DE iMAR Mono+ reduced the artifacts by the greatest amount in the quantitative analysis.
Conclusion:
The method that used DE iMAR Mono+ showed the best performance of spine metal artifact reduction using DECT data. These results may be specific to this CT vendor and implant type.
More Related Videos
10:28Gene Regulation and Targeted Therapy in Gastric Cancer Peritoneal Metastasis: Radiological Findings from Dual Energy CT and PET/CT
Published on: January 22, 2018
08:36The Immersive Cleveland Clinic Virtual Reality Shopping Platform for the Assessment of Instrumental Activities of Daily Living
Published on: July 28, 2022
Related Concept Videos
Alkali Metals
Table 1: Properties of the alkali metals
Bonding in Metals
Oxidation-Reduction Reactions
Metal-Ligand Bonds
In these complexes, transition metals form coordinate covalent bonds, a kind of Lewis acid-base interaction in which both of the electrons in the bond are contributed by a donor (Lewis base) to an electron acceptor (Lewis acid). The Lewis acid in...
Ionization Energy
Properties of Transition Metals