Accuracy and Safety Study of Intracavitary Electrocardiographic Guidance for Peripherally Inserted Central Catheter
Qiying Ling1, Hong Chen, Min Tang
1Department of Pediatrics, Sichuan Academy of Medical Sciences & Sichuan Provincial People's Hospital and Department of Pediatrics, Clinical College of University of Electronic Science and Technology of China (Mr Binzhi Tang and Mss Ling, Chen, and Min Tang, and Mr Binzhi Tang), Chengdu, Sichuan Province, China); and Key Laboratory of Obstetric & Gynecologic and Pediatric Diseases and Birth Defects of Ministry of Education, Sichuan University, Chengdu, Sichuan Province, China (Ms Qu).
Abstract:
The purpose of this study is to investigate the accuracy and safety of intracavitary electrocardiogram (IC-ECG) guidance for the localization of peripherally inserted central catheter (PICC) in neonatal patients. A total of 160 neonatal patients were randomly assigned to receive either anthropometric measurement combined with IC-ECG guidance (n = 80) or conventional anatomical landmark guidance (n = 80) for PICC catheter tip positioning. The catheter tip position was confirmed by postinsertion radiograph and data were interpreted by independent radiologists. Subsequent catheter-related complications of neonates between 2 groups were also compared. The first-attempt target rate was 95.0% (95% confidence interval, 90.1%-99.9%) in IC-ECG-guided PICCs, significantly higher than 78.8% (95% confidence interval, 69.6%-87.9%) in the anatomical landmark guidance group (P < .05). In contrast, IC-ECG-guided PICCs provided a significantly lower overall incidence of the catheter-related complications (3.75%), compared with those guided by anatomical landmarks only (23.75%). Thus, combined use of anatomical landmark and IC-ECG guidance improved the first-attempt target rate of PICC placement and decreased catheter-related complications. These findings indicated a superior accuracy and safety of IC-ECG guidance to conventional anatomical landmark method in neonatal PICC practice.
Related Concept Videos
Peripherally and Centrally Acting Muscle Relaxants: A Comparison
Centrally acting muscle relaxants can be further divided into spasmolytic and antispasmodic drugs. Spasmolytic...
Improving Translational Accuracy
Improving Translational Accuracy
Uncertainty in Measurement: Accuracy and Precision
Survey Safety
The Central Dogma