Conformable Wearable Electrodes: From Fabrication to Electrophysiological Assessment
Published on: July 22, 2022
Validation of wearable textile electrodes for ECG monitoring
Yayoi Tetsuou Tsukada1, Miwa Tokita2, Hiroshige Murata2
1Department of Cardiovascular Medicine, Nippon Medical School, 1-1-5 Sendagi, Bunkyo-Ku, Tokyo, 113-8603, Japan. yayotuka@nms.ac.jp.
Abstract:
A highly conductive textile was woven from nano-fibers coated with the PEDOT-PSS polymer. The aim of this study was to assess the usefulness of textile electrodes for ECG recording as a smart garment. Electrode textile pads and lead wires were sewn to the lining of sportswear and their tolerability to repeated washings were tested up to 150 times. The electrical conductivity of the textile electrode remained functional for up to 50 machine washes. To assess the level of motion artifacts or noise during the daily monitoring of ECG, a single lead ECG with conventional or textile electrodes was recorded during supine rest, seated rest, upright trunk rotation (i.e., twisting), and stepping movement in 66 healthy adults. A Holter system was used for data storage and analysis. ECG patterns of P, QRS, and T waves were comparable between the conventional and textile electrodes. However, the signal-to-artifact-and/or-noise ratio (SAR) during twisting was larger in the textile electrodes than in the conventional electrodes. No skin irritation was seen in the textile electrodes. The single lead textile electrodes embedded in an inner garment were usable for continuous and/or repeated ECG monitoring in daily life except during vigorous trunk movement.
Related Concept Videos
Reliability and Validity
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Standard Electrode Potentials
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Data Validation
Key parameters for method validation include:
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...