Evaluation of different soft tissue-simulating materials in pixel intensity values in cone beam computed tomography
Gustavo Machado Santaella1, Maria Augusta Portella Guedes Visconti2, Karina Lopes Devito3
1Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, State University of Campinas, Piracicaba, São Paulo, Brazil.
Expanded polystyrene (EPS) and utility wax effectively simulated soft tissues for cone beam computed tomography (CBCT) imaging in most regions. These materials offer a viable alternative for simulating soft tissues in dental imaging research.
Area of Science:
- Dental Imaging
- Biomaterials Science
Background:
- Accurate simulation of soft tissues is crucial for interpreting cone beam computed tomography (CBCT) images in dental research.
- Understanding the influence of soft tissues on CBCT image quality is essential for diagnostic accuracy.
Purpose of the Study:
- To evaluate the efficacy of various materials as soft tissue simulators in CBCT.
- To assess the impact of intact soft tissues versus simulated tissues on CBCT imaging outcomes.
Main Methods:
- Piglet heads were imaged with intact soft tissues, stripped tissues, and four different simulated soft tissue materials (acrylic, water, utility wax, expanded polystyrene).
- Pixel intensity values were analyzed for teeth and alveolar bone regions using analysis of variance (ANOVA).
Main Results:
- Expanded polystyrene (EPS) with or without utility wax demonstrated effectiveness in simulating soft tissues for most evaluated regions.
- Acrylic showed moderate effectiveness, while water proved ineffective as a soft tissue simulator.
- Statistical differences were noted in the lower anterior bone and upper anterior tooth regions across various material groups.
Conclusions:
- EPS (2-cm thick), with or without utility wax (1-cm thick), is a suitable material for simulating soft tissues in CBCT imaging for most anatomical areas.
- Acrylic (0.5 cm) is a less effective but viable alternative.
- The influence of intact soft tissues on CBCT image quality was minimal in most regions evaluated.
More Related Videos
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Biodiversity and Human Values
Professional Values
The values that are the foundation of the nursing profession are altruism, autonomy, human dignity, and social justice.
First, altruism refers to the concern for the welfare and well-being of others without personal...
Critical Values
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...