Three Different Protocols of Corneal Collagen Crosslinking in Keratoconus: Conventional, Accelerated and Iontophoresis
Published on: November 12, 2015
Corneal Epithelial Thickness Intrasubject Repeatability and its Relation With Visual Limitation in Keratoconus
Alfredo Vega-Estrada1, Michael Mimouni2, Elias Espla3
1Department of Cornea and Refractive Surgery, Vissum, Alicante, Spain; Keratoconus Unit of Vissum, Alicante, Spain; Division of Ophthalmology, Universidad Miguel Hernández, Alicante, Spain.
Epithelial thickness measurements using AS-OCT and MS-39 are highly repeatable in keratoconus patients. Thinner central and varied peripheral epithelial thickness correlate with visual limitation and keratoconus severity.
Area of Science:
- Ophthalmology
- Corneal Imaging
- Refractive Surgery
Background:
- Keratoconus is a progressive corneal ectasia affecting visual acuity.
- Accurate measurement of corneal epithelial thickness is crucial for diagnosis and management.
- Anterior Segment Optical Coherence Tomography (AS-OCT) offers high-resolution imaging of the cornea.
Purpose of the Study:
- To assess the intra-subject repeatability of epithelial thickness measurements obtained with AS-OCT combined with Placido disc MS-39.
- To correlate epithelial thickness variations with the degree of visual limitation in keratoconus patients.
Main Methods:
- A cross-sectional study involving 107 keratoconus eyes and 60 healthy eyes.
- Three consecutive epithelial thickness measurements were performed using MS-39 in central and peripheral corneal regions.
- Intrasubject repeatability was analyzed using the intraclass correlation coefficient (ICC).
- Correlation between epithelial indices and best-corrected visual acuity (BCVA) was evaluated.
Main Results:
- High repeatability (ICC ≥0.90) was observed for epithelial thickness measurements in both keratoconus and healthy eyes.
- Thinner central 3mm and thicker superior/thinner inferior 8mm epithelial zones were associated with worse BCVA.
- Decreased central 3mm thickness and increased superior-inferior ratios correlated with increased keratoconus severity.
- Multivariate analysis identified thinner central 3mm, superior-inferior ratios, and steeper K2 as significant predictors of keratoconus (AUC=0.92).
Conclusions:
- The MS-39 device demonstrates high repeatability for corneal epithelial thickness measurements.
- Epithelial thickness patterns are significantly correlated with visual acuity and keratoconus grading.
- These AS-OCT-derived epithelial measurements may aid in differentiating keratoconus from healthy eyes.