Performance evaluation of a new and improved cuvette-based automated urinalysis analyzer with phase contrast
Rosanna Falbo1, Maria Roberta Sala1, Marco Bussetti1
1University Department of Laboratory Medicine, Desio Hospital, Via Mazzini 1, Desio, MB, Italy.
Background:
The use of phase contrast in urinalysis has been highly recommended. A new system, sediMAX conTRUST PRO, is now available providing simultaneous automated phase contrast and bright field microscopy. This study aimed to evaluate both analytical and diagnostic performance of this new analyzer.
Methods:
Results from 504 samples evaluated with the sediMAX conTRUST PRO were compared to those obtained from the same samples by manual microscopy (MM). Analytical and diagnostic performance were assessed according to established guidelines.
Results:
The concentration of red blood cells (RBCs)and white blood cells (WBCs) at which the LoQ satisfied a CV< 25% was 12 particles per μL (p/μL) and 8 p/μL, respectively. Within one grade of agreement concordance was quite high, 97.8% for RBCs and 98.0% for WBCs, and above 90% for all other particles. Overall, diagnostic sensitivity and specificity were good (>80%) for the particles considered, although lower sensitivities, 70.6% and 61.8%, were respectively found for hyaline and pathological casts.
Conclusions:
The sediMAX conTRUST PRO provides very good performance in terms of RBC and WBC recognition and enumeration, and quite good performance for all other particles. Hyaline cast and pathological cast identification is fine and comparable to other automated systems, but could use further improvement.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Urine Studies I: Urinalysis
Phase Diagrams
Phase Transitions
Conservative Site-specific Recombination and Phase Variation
The recognition sites for Cre recombinase called LoxP...
Inductance: Single-Phase And Three-Phase Line
Single-Phase Two-Wire Line:
A single-phase line consists of two solid cylindrical conductors, denoted as x and y. Each conductor carries phasor currents ix and iy, respectively. Given that the sum of these currents is...