Novel adhesives for sternal fixation and stabilization: A biomechanical analysis
Cina Mehrvar1, Paul Kuzyk2, Gideon Cohen3
1Department of Mechanical and Industrial Engineering, Ryerson University, 350 Victoria St, Toronto, ON M5B 2K3, Canada.
Background:
Cerclage wires remain the current standard of care following median sternotomy, despite significant complications including dehiscence and infection. This study uses a human cadaveric model to investigate the use of glass polyalkenoate cements formulated from two glasses, A (mole fraction: SiO2:0.48, ZnO:0.36, CaO:0.12, SrO:0.04) and B (mole fraction: SiO2:0.48, ZnO:0.355, CaO:0.06, SrO:0.08, P2O5:0.02, Ta2O5:0.005), to improve wired sternal fixation.
Methods:
Median sternotomies were performed on fifteen cadaveric sterna. Fixation was performed with either traditional wire cerclage or adhesive-enhanced wire cerclage; the adhesive based on either Glass A or Glass B. Cyclic tensile loading of 10 N to 100 N was applied. Every 30 cycles, the maximum load was increased by 100 N up to a maximum of 500 N. Two adhered sterna were tested beyond 500 N. Mid-sternal displacement was measured to assess fixation stability.
Findings:
Displacement for adhesive-enhanced sternal closures were significantly less (p < 0.05) than standard wire cerclage. There was no significant difference between adhesives. Up to 500 N, no adhesive-enhanced sternum experienced a pathological sternal displacement (>2 mm), while three out of five of traditional wire fixations did. Of the two adhered samples tested beyond 500 N, one showed pathological displacement at 800 N and the other at 1100 N. Failure of adhered sterna appeared to initiate within the trabecular bone rather than in the adhesive.
Interpretation:
The adhesives were capable of providing immediate bone stability, significantly reducing sternal displacement. In vivo investigations are warranted to determine the effect the adhesives have on bone remodelling.
Related Concept Videos
Adhesion
Capillary action is a result of water’s adhesive tendencies. When a narrow...
Nuclear Stability
To hold positively charged protons together...
RNA Stability
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
Stability
The stability of an LTI system is determined by the roots of its characteristic equation, known as poles. A system is stable if it produces a bounded...
Carbon-dioxide Fixation