Objectives:
Measure the sound levels produced by various lasers commonly used during routine outpatient cosmetic surgery to determine whether or not their use exceeds exposure levels set forth by the US National Institute for Occupational Safety and Health (NIOSH) and the US Occupational Safety and Health Administration (OSHA).
Materials And Methods:
Using two different meters, the sound levels of lasers commonly used in cosmetic surgery were recorded during various procedures for several indications: tattoo removal, treatment of lentigines and pigmented lesions, facial erythema and vascular lesions, hair removal and resurfacing of acne scars, and photoaging.
Results:
All but five lasers had a maximum sound level below 85 dBA, the limit proposed by NIOSH. The loudest laser examined was a fully ablative 2,940 nm Er:YAG during facial resurfacing, with an average maximum sound level of 101.5 decibels (dBA). Two other lasers used for resurfacing exceeded 85 dBA including a fractional ablative 1064 Nd:YAG with an average maximum of 97.8 and a different fully ablative 2,940 nm Er:YAG which had an average maximum of 96.3 nm. The two other lasers that exceeded 85 dBA were picosecond lasers used to treat black tattoos, including a 1,064 nm Nd:YAG with an average maximum of 93.7 dBA and a 755 nm alexandrite with an average maximum of 93.6 dBA.
Conclusion:
Although some lasers in cosmetic surgery may be perceived as being quite loud, they remain safe. Even the loudest laser studied would have to be used for nearly 2 hours before exceeding the OSHA recommended exposure limit. Even physicians who spend a large amount of time using lasers in clinical practice should be reassured that these devices are not likely to produce noise-induced hearing loss (NIHL) hearing loss. Lasers Surg. Med. 51:491-494, 2019. © 2019 Wiley Periodicals, Inc.
More Related Videos
07:05Author Spotlight: Eco-Friendly Extraction of Bioactive Compounds Using Polyol-Based Microwave-Assisted Techniques
Published on: August 23, 2024
11:46Investigation of the Electrophysiological and Thermographic Safety Parameters of Surgical Energy Devices During Thyroid and Parathyroid Surgery in a Porcine Model
Published on: October 13, 2022
Related Concept Videos
Sound Intensity Level
The human ear can perceive an extensive range of sound intensity, necessitating the use of the logarithmic scale to define a physical quantity—the intensity level. It is a ratio of two intensities and...
Survey Safety
Korotkoff Sounds
During blood pressure assessment, inflating the cuff 30 millimeters of mercury above the patient's systolic blood pressure...
Heart Sounds
Auscultation is the process of listening to these internal body sounds using a stethoscope. The heart produces four types of sounds, but only two—S1 and S2—can usually be heard with a stethoscope.
S1, also known as the "lub" sound, is caused by the closure of atrioventricular (A-V)...
Soundness of Cement
Sound Waves
Sound waves are longitudinal in most fluids because fluids cannot sustain any lateral pressure. In solids, however, shear forces help in propagating the disturbance in the lateral direction as well....