Related Experiment Video
Updated: Jan 29, 2026

Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
Utility of point-of-care vs reference laboratory testing for the evaluation of glucose levels
O M Andriankaja1, F J Muñoz-Torres1, J L Vergara1
1Center for Clinical Research and Health Promotion, School of Dental Medicine, University of Puerto Rico Medical Sciences Campus, San Juan, Puerto Rico.
Aims:
To assess the level of agreement between point-of-care and laboratory reference glucose values in defining glycaemic status.
Methods:
We analysed 1292 overweight/obese, non-institutionalized participants, aged 40-65 years, in the San Juan Overweight Adults Longitudinal Study. Fasting venous blood glucose was determined using a point-of-care Bayer Contour Blood Glucose Meter and by Vitros System 250 instrument (laboratory). American Diabetes Association thresholds were used to classify participants into normoglycaemia (< 5.6 mmol/l), prediabetes (5.6 to 6.9 mmol/l), or diabetes groups (≥ 7 mmol/l).
Results:
Bland-Altman plot analysis showed a slope of 0.04 (P=0.002) for the regression between the mean difference and the average of the two methods. The slopes were significantly different from zero among people with normoglycaemia (β=-0.57, P<0.001), and prediabetes (β=-0.75, P<0.001) but not among people with diabetes (β=-0.02, P=0.68). When the prediabetes and diabetes groups were merged into one group, the slope was 0.01, and the glucose values remained similar using the two methods (P=0.76).
Conclusion:
Point-of-care blood glucose measurement may be useful to screen people with diabetes, and to assess glucose among individuals with diabetes where blood can be drawn, but laboratory tests are unavailable or untimely.
More Related Videos
Related Concept Videos
Traditional Level Of Health Care System
The preventive healthcare service includes tests for screening. Preventive health care services include identifying and reducing disease risk...
Inertial Frames of Reference
Non-inertial Frames of Reference
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
Glucose Transporters
Facilitated diffusion-glucose transporters (GLUTs) are encoded by the solute-linked carrier (SLC) family 2, subfamily A gene family, or SLC2A. The 14 GLUT protein members are distributed into three classes:
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...