Higher-order aberration measurements: Comparison between Scheimpflug and dual Scheimpflug-Placido technology in
Andre L Piccinini1, Oren Golan2, Farhad Hafezi3
1Keck School of Medicine, University of Southern California, Los Angeles, USA; Sadalla Amin Ghanem Eye Hospital, Joinville, SC, Brazil.
Purpose:
To compare higher order aberrations (HOAs) in normal eyes between a Scheimpflug imaging system (Pentacam HR) and dual Scheimpflug-Placido imaging system (Galilei G4).
Setting:
Emory University, Atlanta, Georgia, and the University of Southern California, Los Angeles, USA.
Design:
Retrospective case series.
Methods:
Eyes screened for refractive surgery were evaluated sequentially using a Scheimpflug device and a dual Scheimpflug-Placido device. Differences, correlations, and agreement between values for total root mean square (RMS), trefoil, coma, and spherical aberration were analyzed, and Bland-Altman plots were generated.
Results:
One hundred five eyes of 105 patients (44 men, 61 women) were evaluated. Total RMS, coma, and trefoil were significantly different between groups (all P < .001), while spherical aberration values were not. There was moderate correlation between devices for trefoil (r = 0.475 to 0.652), coma (r = 0.574 to 0.651), and spherical aberration (r = 0.483) and a strong correlation for total cornea RMS (r = 0.817). There was no directional bias between groups. The 95% limits of agreement for absolute values was 0.039 μm for trefoil at 30 degrees, 0.405 μm for trefoil at 0 degree, 0.553 μm for horizontal coma, 0.545 μm for vertical coma, 0.318 μm for spherical aberration, and 0.617 μm for RMS.
Conclusions:
A Scheimpflug imaging device and dual Scheimpflug-Placido imaging device generated statistically different values for total cornea HOAs; however, the correlation between devices was moderate to strong and there was reasonable agreement in all measures for normal eyes. Based on these findings, the devices appear functionally equivalent for clinical use, although caution is warranted for outcomes-based research protocols that report HOAs.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Normal Stress
When a rod is under axial loading, the internal forces and corresponding stress are normal to the plane of the section, so it is termed normal stress. It's important to...
Normal Distribution
Variation: Normal Distribution, Range, and Standard Deviation
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Normal and Shear Force