A Coaxial RF Applicator for Ultra-High Field Human MRI
Objective:
To develop a novel radio-frequency (RF) concept for ultra-high field (UHF) human magnetic resonance imaging (MRI) based on a coaxial resonant cavity.
Methods:
A two-channel slotted coaxial cavity RF applicator was designed for human head MRI at 9.4T. Physical dimensions made the proposed conducting structure resonant at the required frequency without tuning lumped elements. Numerical electromagnetic modeling was used to optimize the design. RF safety was assessed with two representative human body models. MR experiments on a 9.4T scanner included gradient echo images and mapping of a circularly polarized RF magnetic field in the human head phantom.
Results:
The simulations and the phantom MR experiments agreed both qualitatively and quantitatively. The design was relatively simple, robust and required only a few additional reactive elements for the applicator's input impedance matching. The transmit efficiency and homogeneity of the excitation field were only 20% and 4% lower compared to a conventional 8-channel head array.
Conclusion:
The coaxial RF applicator was feasible for human MRI at UHF and required no lumped elements for its tuning. Imaging performance of the RF applicator was only moderately lower compared to the conventional transmit array, but would be sufficient to provide an anatomical reference for the heteronuclei MRI.
Significance:
An alternative approach with the minimal involvement of lumped elements becomes feasible to design volume-type RF coils for UHF human MRI.
Related Concept Videos
Induced Electric Fields: Applications
Field Application of Global Positioning System
Energy Stored In A Coaxial Cable
In the simplest form, a coaxial cable can be represented by two long hollow concentric cylinders in which the current flows in opposite directions. The magnetic field inside and outside the coaxial cable is determined by using Ampère's law. The magnetic field inside...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Field Effect Transistor
Electric Field
In the new picture, imagine that the first charge sets up an electric field independent of all other charges in the universe. When another charge comes in its vicinity, the second charge experiences an electric force depending on the electric field at that point. The source charge does not...