[Comparison of ultrasound and hydrodynamic parameters in phacoemulsification]
S V Shukhaev1, O M Yeltsina2, L I Balashevich1
1Saint Petersburg branch of S. Fyodorov Eye Microsurgery Federal State Institution, 21 Yaroslava Gasheka St., Saint Petersburg, Russian Federation, 192283.
Abstract:
Modern phacoemulsifiers have the widest selection of ultrasound parameters. Existing approaches to comparing the settings of the phaco machines have shortcomings and thus cannot be considered objective.
Purpose:
To find a technique that would allow objective in vivo evaluation of various ultrasound and hydrodynamic parameters after performing phacoemulsification. To compare the two types of phaco settings using the proposed technique.
Material And Methods:
The principle of the 'Femtocomparison' method is to have each patient undergo femtolaser pretreatment before phacoemulsification, in which the lens is divided by laser into equal parts and a corneal incision and capsulotomy are performed with the same settings. Then, during the manual part, basic ultrasound and hydrodynamic data is recorded after the removal of each quadrant of the lens. The study included 58 patients with cataracts of varying densities who underwent femtosecond laser-assisted cataract surgery. In each case, one half of the nucleus was removed using torsional ultrasound with IP technology, while the other half - using a combination of torsional and longitudinal ultrasound.
Results:
The 'Femtocomparison' technique allowed objective in vivo comparison of two types of ultrasound settings. When assessing these parameters, no significant differences were found in any of the patients regardless of the density of the lens nucleus (p>0.05). In soft nuclei group (NC 1-3) CDE were significantly lower when using torsional ultrasound and IP (p<0.05). In the group with dense nuclei (NC 6+), CDE and aspiration time were significantly lower when using combination ultrasound (p<0.05).
Conclusion:
The 'Femtocomparison' technique has shown that torsional ultrasound with IP technology is preferable for aspiration of soft cataracts, while combined use of torsional and longitudinal ultrasound is better when removing dense lenses.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Wave Parameters
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...