Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Classification of Connective Tissues01:30

Classification of Connective Tissues

15.8K
The connective tissues have different properties and functions in the human body. They are broadly categorized into proper, supporting, or fluid connective tissues.
Connective Tissue Proper
Connective tissue proper is the most abundant class of connective tissues. As its name implies, it predominantly connects different tissues in the body. Depending on the cell types, ground substance, viscosity, and fiber types in the ECM, connective tissue proper is further categorized into loose and dense....
15.8K
Classification of Epithelial Tissues: Overview01:22

Classification of Epithelial Tissues: Overview

21.8K
Epithelial tissues are classified according to the shape of the cells and the number of cell layers formed. Cell shapes can be squamous (flattened and thin), cuboidal (square-like, as wide as it is tall), or columnar (rectangular, taller than it is wide). Additionally, the nucleus shape helps identify the type of epithelial cells. Squamous cells have flattened disc-shaped nuclei, cuboidal cells have spherical nuclei, and columnar cells have elongated nuclei.
Based on the number of cell layers,...
21.8K
Classification of Epithelial Tissues: Simple Epithelium01:30

Classification of Epithelial Tissues: Simple Epithelium

13.1K
Simple epithelium consists of a single layer of cells that lines body cavities and blood vessels. The shape of the cells in the epithelium reflects the function of the tissue. Cells in simple squamous epithelium appear as thin scales with flat, elliptical nuclei that mirror the form of the cell.
Because of the thinness of the cells, simple squamous epithelium is present where the rapid passage of chemical compounds is observed. For example, the endothelium that lines the capillaries and vessels...
13.1K
Classification of Epithelial Tissues: Stratified Epithelium01:29

Classification of Epithelial Tissues: Stratified Epithelium

12.7K
Stratified epithelium consists of several stacked layers of cells. They provide the durability to withstand constant physical and chemical attacks. Stratified epithelium is named after the shape of the most apical layer of cells. Stratified squamous epithelium is the most common type found in the human body. In this tissue, the apical cells are squamous, whereas the basal layer contains either columnar or cuboidal cells. The basal cells divide to form new daughter cells, which gradually become...
12.7K
Classification of Epithelial Tissues: Glandular Epithelium01:20

Classification of Epithelial Tissues: Glandular Epithelium

11.9K
The glandular epithelium is made of one or more epithelial cells modified to synthesize and secrete chemical substances. Glandular epithelia can be classified based on cell number. Unicellular glands have individual secretory cells scattered across the epithelial monolayer. In contrast, multicellular glands consist of a hollow tubular duct attached to the cluster of secretory cells located in the deep pockets.
Multicellular glands are formed during early development when epithelial budding...
11.9K
Distance Corrections01:15

Distance Corrections

284
To achieve precise distance measurements, especially in surveying and construction, certain corrections must be applied to account for potential sources of error like the standardization errors, temperature variations, and slope adjustments.Standardization error emerges when measurement equipment undergoes changes, such as wear, repairs, or weather impacts. To address this, surveyors compare the equipment’s readings to a standard. This process identifies any deviation that might lead to...
284

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Nonstop gated CBCT for respiratory gating lung SBRT: A feasibility study.

Medical physics·2025
Same author

PFCM: Poisson Flow Consistency Models for Low-Dose CT Image Denoising.

IEEE transactions on medical imaging·2025
Same author

One High-Volume Posteromedial Knee Capsular Injection Can Provide Safe Anesthetic Infiltration Across the Entire Posterior Knee Capsule.

Arthroplasty today·2025
Same author

Coronary atherosclerotic plaque characterization with silicon-based photon-counting computed tomography (CT): A simulation-based feasibility study.

Medical physics·2024
Same author

Low-dose computed tomography perceptual image quality assessment.

Medical image analysis·2024
Same author

Dual-Energy CTA Iodine Map Reconstructions Improve Visualization of Residual Cerebral Aneurysms following Endovascular Coiling.

AJNR. American journal of neuroradiology·2024

Related Experiment Video

Updated: Jan 29, 2026

Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
09:49

Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization

Published on: December 2, 2013

10.8K

Fast shading correction for cone-beam CT via partitioned tissue classification.

Linxi Shi1,2, Adam Wang1, Jikun Wei3,4

  • 1Department of Radiology, Stanford University, Palo Alto, CA 94305, United States of America.

Physics in Medicine and Biology
|February 6, 2019
PubMed
Summary

This study introduces a new method to correct shading artifacts in cone beam computed tomography (CBCT) images for radiation therapy without needing prior CT scans. The technique significantly improves image quality, making CBCT more reliable for clinical use.

More Related Videos

Single-Digit Nanometer Electron-Beam Lithography with an Aberration-Corrected Scanning Transmission Electron Microscope
10:25

Single-Digit Nanometer Electron-Beam Lithography with an Aberration-Corrected Scanning Transmission Electron Microscope

Published on: September 14, 2018

10.6K
SpOT the Correct Tissue Every Time in Multi-tissue Blocks
06:53

SpOT the Correct Tissue Every Time in Multi-tissue Blocks

Published on: May 31, 2015

12.3K

Related Experiment Videos

Last Updated: Jan 29, 2026

Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
09:49

Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization

Published on: December 2, 2013

10.8K
Single-Digit Nanometer Electron-Beam Lithography with an Aberration-Corrected Scanning Transmission Electron Microscope
10:25

Single-Digit Nanometer Electron-Beam Lithography with an Aberration-Corrected Scanning Transmission Electron Microscope

Published on: September 14, 2018

10.6K
SpOT the Correct Tissue Every Time in Multi-tissue Blocks
06:53

SpOT the Correct Tissue Every Time in Multi-tissue Blocks

Published on: May 31, 2015

12.3K

Area of Science:

  • Medical Imaging
  • Radiation Oncology
  • Image Processing

Background:

  • Cone beam computed tomography (CBCT) is vital in radiation therapy but suffers from shading artifacts, limiting its quantitative use.
  • Existing correction methods often require planning CT (pCT) images, reducing clinical practicality.

Purpose of the Study:

  • To develop an effective CBCT shading artifact correction method that does not require prior CT images.
  • To enhance the clinical applicability and image quality of CBCT for radiation therapy.

Main Methods:

  • A novel method utilizing partitioned tissue classification to generate a 'shading-free' template image from CBCT data.
  • Employing sparse sampling of shading artifacts and local filtration (a Fourier transform-based algorithm) for efficient correction.
  • Validation using an anthropomorphic pelvis phantom and 6 pelvis patient datasets.

Main Results:

  • The proposed method significantly reduces signal non-uniformity (SNU) and maximum CT number errors in CBCT images of phantoms and patients.
  • On average, SNU was reduced from 9.22% to 1.06% (fat) and 11.41% to 1.67% (muscle) in patient data.
  • Maximum CT number errors were reduced from 95 HU to 9 HU (fat) and 88 HU to 8 HU (muscle) in patient data.
  • Processing time for one CBCT dataset is approximately 45 seconds on a standard PC.

Conclusions:

  • The developed method effectively corrects CBCT shading artifacts without prior CT information, achieving image quality comparable to pCT.
  • This approach enhances the quantitative accuracy and clinical utility of CBCT in radiation therapy.
  • The method is efficient and practical for routine clinical application.