Optimal wavelength for the clot waveform analysis: Determination of the best resolution with minimal interference of
Jonathan Evrard1, Romain Siriez1, Laure Morimont1
1Department of Pharmacy, Namur Thrombosis and Hemostasis Center (NTHC), Namur Research Institute for Life Sciences (NARILIS), University of Namur, Namur, Belgium.
Introduction:
Clot waveform analysis (CWA), a new methodology to assess coagulation process, can be usefully applied in various clinical settings. However, its clinical use is limited mainly because of the absence of standardization. No consensus exists regarding the wavelengths at which CWA has to be performed what is crucial for the sensitivity of the CWA.
Objectives:
The primary aim of this study is to determine which wavelength is the most sensitive and specific for CWA. Interindividual baseline absorbance will also be assessed as the impact of reagents from the intrinsic, extrinsic, and common coagulation pathway will be determined.
Methods:
Plasma samples were screened at wavelengths from 280 to 700 nm to provide absorbance spectra in clotted and nonclotted plasma. The interindividual variability of baseline absorbance was obtained by screening plasma from 50 healthy individuals at 340, 635, and 671 nm. The inner-filter effect of reagents was assessed in plasma or serum when appropriate at the same wavelengths. The reagents were those commonly used for activated partial thromboplastin time, prothrombin time, thrombin time, and dilute Russell's viper venom time.
Results:
Clotted plasma has higher absorbance value than nonclotted plasma (P < 0.01). The absorbance of all type of samples is higher at 340 nm than at >600 nm (P < 0.01). The interindividual variability at the different wavelengths was around 25%. However, except with the STA®-CKPrest® and STA®-NeoPTimal®, the reagents do not have a significant effect on the baseline absorbance.
Conclusions:
Wavelengths above 650 nm are recommended to perform CWA. Most of the commercialized reagents can be used for CWA.
More Related Videos
Related Concept Videos
The de Broglie Wavelength
Interference and Diffraction
RNA Interference
This process occurs naturally in cells, often through the activity of genomically-encoded microRNAs. Researchers can take advantage of this mechanism by introducing synthetic RNAs to deactivate specific genes for research or therapeutic purposes. For example, RNAi could be used...
Effective Value of a Periodic Waveform
The effective value of a periodic current represents the direct current (DC) that conveys the same average power to a resistor as the periodic current itself. This concept is crucial when assessing AC circuits. To determine the...
Clot Retraction and Fibrinolysis
EDTA: Auxiliary Complexing Reagents