Extended Protective Shield Under Table to Reduce Operator Radiation Dose in Percutaneous Coronary Procedures
Alessandro Sciahbasi1, Alessandro Sarandrea2, Stefano Rigattieri1
1Interventional Cardiology, Sandro Pertini Hospital, Rome, Italy (A. Sciahbasi, S.R., R.P., M.C., C.D.R., S.F., G.F.).
Adjunctive under-the-table radiation shields significantly reduce operator radiation dose at the pelvic and thorax levels during percutaneous coronary procedures. This study confirms their effectiveness in minimizing exposure for medical professionals.
Area of Science:
- Interventional Cardiology
- Radiation Safety
- Medical Imaging
Background:
- Operator radiation exposure remains a concern during percutaneous coronary procedures, particularly at the pelvic region.
- Existing protective tools effectively reduce thoracic radiation dose but not pelvic dose.
Purpose of the Study:
- To evaluate the efficacy of under-the-table adjunctive shields in reducing operator radiation exposure during percutaneous coronary procedures.
Main Methods:
- A prospective, single-center, randomized study (EXTRA-RAD) involving 205 transradial coronary procedures.
- Patients were randomized into two groups: standard arrangement and adjunctive under-the-table shield.
- Two types of under-the-table shields were compared within the shield group.
Main Results:
- Adjunctive shields significantly reduced radiation dose at the pelvic region (42 µSv vs. 13 µSv, P<0.0001).
- Shields also lowered radiation dose at the thorax level (4 µSv vs. 2 µSv, P=0.001).
- Dose reduction was consistent across all operators, with no significant difference between shield types.
Conclusions:
- Under-the-table adjunctive shields are effective in significantly lowering operator radiation dose at both pelvic and thoracic levels during transradial coronary procedures.
- These shields represent a valuable tool for enhancing radiation safety in interventional cardiology.