A Comparative Study of SEER Sonorheometry Versus Standard Coagulation Tests, Rotational Thromboelastometry, and
Ekaterina Baryshnikova1, Umberto Di Dedda1, Marco Ranucci1
1Department of Cardiovascular Anaesthesia and Intensive Care Unit, IRCCS Policlinico San Donato, San Donato Milanese (Milan), Milan, Italy.
Objectives:
Coagulation point-of-care tests for the diagnosis and management of bleeding commonly are used in cardiac surgery. Recently, a new viscoelastic point-of-care device, Quantra (HemoSonics, Charlottesville, VA), which is based on sonic estimation of elasticity via resonance sonorheometry, entered the market. Prior studies have compared the Quantra parameters with conventional coagulation tests and thromboelastography or rotational thromboelastography, but no study has assessed the correlation of the platelet-derived Quantra parameter platelet contribution to stiffness (PCS) with platelet function assays. The present study compared the Quantra-derived coagulation parameters with conventional coagulation tests, rotational thromboelastography-derived parameters, and platelet function measured using multiplate aggregometry.
Design:
Prospective cohort study.
Setting:
University research hospital.
Participants:
The study comprised 30 cardiac surgery patients before and after cardiopulmonary bypass.
Interventions:
None.
Measurements And Main Results:
Quantra clot stiffness showed a very strong correlation with the EXTEM maximum clot firmness before and after cardiopulmonary bypass and in the overall samples (r values 0.94-0.96). Quantra fibrinogen contribution to stiffness was compared with the corresponding FIBTEM and Clauss fibrinogen levels, which also showed a strong level of correlation (r values between 0.74 and 0.87). Quantra PCS strongly correlated with platelet count (r = 0.71) in the overall samples and moderately with adenosine diphosphate-dependent platelet function (r = 0.67). In a multivariable model, both the adenosine diphosphate test value and the platelet count remained independently associated with Quantra PCS.
Conclusions:
Fibrinogen-dependent clot stiffness properties are well-reflected by the Quantra fibrinogen contribution to stiffness parameter, and PCS incorporates platelet count and function.
More Related Videos
Related Concept Videos
Standard Electrode Potentials
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Testing a Claim about Standard Deviation
The hypothesis testing for the claim of population standard deviation (or variance) requires the data and samples to be random and unbiased. The population distribution also must be normal. There is no specific requirement on the sample size as the estimation is based on the chi-square distribution.
As a first step, the hypothesis (null and alternative) concerning the claim about...
Coagulation
Coagulation
During the coagulation phase, clotting factors, or procoagulants, play a vital role in initiating and progressing the coagulation cascade. This cascade is a series of reactions...
Bioavailability Study Design: Single Versus Multiple Dose Studies