Biomechanical testing of different posterior fusion devices on lumbar spinal range of motion
Frank Hartmann1, Thomas Nusselt1, Gerrit Maier2
1Center for Trauma and Orthopedic Surgery, Gemeinschaftsklinikum Mittelrhein, Ev. Stift Koblenz, Germany.
Background:
Recent minimal-invasive posterior fusion devices are supposed to provide stability and obtain fusion in combination with interbody cages in the instrumented segment. The aim of the present study is to evaluate the primary stability of two minimal-invasive posterior prototypes compared to an established spinous process plate and standard pedicle screw instrumentation.
Methods:
Seven fresh frozen human cadaver lumbar spines (L2-L5) were tested in a spinal testing device with a moment of 7.5 Nm. Spinal stability was determined as mean range of motion (RoM) in the segment L3/L4 during extension-flexion, lateral bending and axial rotation. The RoM was measured during five conditions: 1. intact; 2. working prototype composed of an interspinous device and process plates; 3. an established spinous process fixation device 4. working prototype of facet fixation and 5. pedicle screw fixation.
Findings:
All devices caused a significant reduction of RoM during extension-flexion. The RoM during lateral bending was significantly reduced to 37% (of intact) by pedicle screws and 68% by facet fixation prototype. During axial rotation the RoM was significantly reduced to 52% by pedicle screws and to 86% by facet fixation prototype. The other devices had no significant influence on RoM during lateral bending and axial rotation.
Interpretation:
The facet fixation prototype provided less primary stability compared to pedicle screws, but had significant advantages over spinous process fixation techniques. The results encourage further testing of this implant as a minimal-invasive approach for posterior fixation.
Related Concept Videos
Nuclear Fusion
A helium nucleus has a mass that is 0.7% less than that of four hydrogen nuclei; this lost mass is converted into energy during the fusion. This reaction produces about...
Range
15.9; 16.1; 15.2; 14.8; 15.8; 15.9; 16.0; 15.5
Measurements of the amount of soda in a 16-ounce can vary since different subjects record these measurements or since the exact amount - 16 ounces of liquid, was not...
¹H NMR: Long-Range Coupling
In alkenes, spin information is communicated via σ–π overlap, as seen in allylic (four-bond) and homoallylic (five-bond) couplings. These coupling interactions are stronger when the σ bond is parallel to the alkene...
Equation of Motion: General Plane motion
Moreover, the body's center of mass experiences a rotational effect as a result of these couple moments. This rotation can be articulated as the...
Variation: Normal Distribution, Range, and Standard Deviation
Angle of Twist - Elastic Range