Enhancing Electrode Location Assessment in Cochlear Implantation via Computed Tomography Image Fusion
Published on: January 17, 2025
Comparison of Skull Radiograph and Computed Tomography Measurements of Cochlear Implant Insertion Angles
Sara Gallant1, David R Friedmann1, Mari Hagiwara2
1NYU School of Medicine, Department of Otolaryngology.
Background:
Measurement of the angular depth of insertion (aDOI) of cochlear implant electrode arrays has numerous clinical and research applications. Plain-film radiographs are easily obtained intraoperatively and have been described as a means to calculate aDOI. CT imaging with 3D reformatting can also be used for this measurement, but is less conveniently obtained and requires higher radiation doses, a particular concern in pediatrics. The extent to which plain-film and 3D CT image-based measurements are representative of the true position of the electrode within the cochlea is unknown.
Methods:
Cochlear implantation was performed on 10 cadaveric temporal bones. Five bones were implanted with perimodiolar electrodes (Contour Advance TM, Cochlear, Sydney, Australia) and five were implanted with lateral wall electrodes (Slim Straight, Cochlear). The insertion depths of the electrodes were varied. Each bone was imaged with a radiograph and CT. aDOI was measured for each bone in each imaging modality by a neurotologist and a neuroradiologist. To obtain a 'gold standard' estimate of aDOI, the implanted temporal bones were embedded in an epoxy resin and methodically sectioned at 100 μm intervals; histologic images were captured at each interval. A 3D stack of the images was compounded, and a MATLAB script used to calculate aDOI of the most apical electrode. Measurements in the three modalities (radiograph, CT, and histology) were then compared.
Results:
The average aDOI across all bones was similar for all modalities: 423° for radiographs, 425° for CT scans, and 427° for histology, indicating that neither imaging modality resulted in large systematic errors. Using the histology-measured angles as a reference, the average error for CT-based measures (regardless of whether the error was in the positive or negative direction) was 12°, and that for radiograph-based measures was 15°. This small difference (12 vs 15° error) was not statistically significant.
Conclusion:
Based on this cadaveric temporal bone model, both radiographs and CTs can provide reasonably accurate aDOI measurements. In this small sample, and as expected, the CT-based estimates were more accurate than the radiograph-based measurements. However, the difference was small and not statistically significant. Thus, the use of plain radiographs to calculate aDOI seems judicious whenever it is desired to prevent unnecessary radiation exposure and expense.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Design Example: Traverse Angle Computations
Imaging Studies III: Computed Tomography
Sutures of the Skull
Sutures are immobile joints between adjacent bones of the skull. The narrow gap between the bones is filled with dense, fibrous connective tissue that unites the bones. The long sutures located between the skull bones are not straight but instead follow irregular, tightly twisting paths. These twisting lines tightly...
Overview of the Skull
The cranial vault surrounds and protects the brain and houses the middle and inner ear structures. This cavity is bounded superiorly by the rounded top of the skull, which...
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...