Comprehensive Comparison of Axial Length Measurement With Three Swept-Source OCT-Based Biometers and Partial
Purpose:
To compare axial length measurements (and failure rate) of three swept-source optical coherence tomography (SS-OCT)-based biometers: IOLMaster 700 (Carl Zeiss Meditec, Jena, Germany), OA-2000 (Tomey, Nagoya, Japan), and Argos (Movu Inc., Komaki, Japan) to those provided by a partial coherence interferometry (PCI)-based optical biometer (IOLMaster v5.4 [Carl Zeiss Meditec]).
Methods:
A total of 119 patients (171 eyes) undergoing cataract surgery were enrolled. Axial length was measured with the four biometers in a random order. Chi-square analysis was used to determine whether statistically significant differences in success rates were found between biometers. Within-subject standard deviation (Sw), test-retest repeatability (TRT), coefficient of variation (CoV), and intraclass correlation coefficients (ICCs) were calculated to analyze the intraobserver repeatability. Bland-Altman plots were used for agreement assessment.
Results:
Axial length measurements were successfully measured in 166 eyes (97.08%) with the IOLMaster 700, 166 eyes (97.08%) with the OA-2000, 170 eyes (99.42%) with the Argos, and 138 eyes (80.70%) with the IOLMaster v5.4. Chi-square analysis indicated a significant difference in the failure rate between PCI- and SS-OCT-based biometers (P < .001). Intraobserver repeatability for the IOLMaster 700, OA-2000, and Argos showed excellent repeatability with low TRT (0.03, 0.06, and 0.05 mm, respectively), low CoV (0.04%, 0.10%, 0.07%, respectively), and high ICC (1.000, 0.999, and 1.000, respectively). The Bland-Altman 95% limits of agreement were as narrow as 0.09 mm, indicating excellent agreement among the three SS-OCT biometers and the PCI biometer.
Conclusions:
The three SS-OCT biometers showed a significantly higher success rate for axial length measurement than the IOLMaster v5.4 in various lens opacities. These SS-OCT biometers are likely to become the gold standard for axial measurement. [J Refract Surg. 2019;35(2):115-120.].
More Related Videos
03:35Determining Gender-Based Differences in Retinal and Choroidal Thickness in Underweight Individuals via Swept-Source Optical Coherence Tomography
Published on: December 1, 2023
10:39Measurement of X-ray Beam Coherence along Multiple Directions Using 2-D Checkerboard Phase Grating
Published on: October 11, 2016
Related Concept Videos
Sources of Self-Esteem III: Social Comparison
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Axial and Appendicular Muscles
Axial Muscles
Axial muscles, situated along the body's midline, are intricately connected to the axial skeleton, which includes the skull, spine, ribs, and sternum. These muscles facilitate facial expressions and...
Overview of the Axial Skeleton
The axial skeleton of the...
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical...
Partial Fractions