Novel external quality assurance scheme for drug susceptibility testing of non-tuberculous mycobacteria: a
Vladyslav Nikolayevskyy1, Florian P Maurer2, Yen Holicka1
1National Mycobacterium Reference Service South, Public Health England, London, UK.
Objectives:
Our aim was to conduct a multicentre study involving laboratories participating in the European TB Reference Laboratory Network aiming to develop a pilot external quality assessment (EQA) scheme for drug susceptibility testing (DST) of non-tuberculous mycobacteria (NTM).
Methods:
The study comprised a survey using a structured questionnaire followed by a pilot EQA round using identical panels of 10 Mycobacterium avium (MAV) and Mycobacterium abscessus (MAB) isolates. EQA results were received from 16 laboratories utilizing the broth microdilution method. Consensus modal MIC values were determined, and essential and categorical agreement rates were calculated.
Results:
Twenty-four out of 31 laboratories (77.4%) reported DST for NTM routinely. Essential agreement ranged from 78.8% (amikacin) to 96.2% (linezolid) for MAV and from 76.0% (amikacin) to 100% (doxycycline) for MAB. Categorical agreement ranged from 56.8% (moxifloxacin) to 100% (clarithromycin) for MAV and from 53.6% (linezolid) to 100% (doxycycline) for MAB.
Conclusions:
Our results show that interlaboratory reproducibility of DST for NTM is insufficient, highlighting the need for expanding EQA schemes. As EQAs for Mycobacterium tuberculosis complex have led to more reliable and reproducible DST, we propose to follow a similar approach for clinically relevant NTM.
Related Concept Videos
Quality Assurance
Testing Water Quality
The Z-Scheme of Electron Transport in Photosynthesis
Pilot and Numeric Relaying
Work Done on a System by External Force
In the presence of a non-conservative opposing force, like friction, some part of the work done...
Magnetic Susceptibility and Permeability
When diamagnetic materials are placed under an external magnetic field, the moments opposite to the field are induced. Hence, the susceptibility for diamagnets has a minimal negative value of 10-5–10-6. Since...