Use of a Percutaneous Ventricular Assist Device/Left Atrium to Femoral Artery Bypass System for Cardiogenic Shock
Published on: August 16, 2021
Reproducibility of Semi-automated Three-dimensional Volumetric Analysis using Cardiac Computed Tomography in Patients
Gauravpal S Gill1, Gaby Weissman1, Yael F Meirovich2
1MedStar Washington Hospital Center, Washington, D.C., United States of America.
Background:
Multi-detector gated cardiac computed tomography (CCT) allows three-dimensional (3D) quantification of cardiac chambers and is clinically indicated to assess left ventricular assist device (LVAD) malfunction and complications. Automated volumetric analysis is, however, disrupted by inflow cannula artifact in patients with LVAD. With this study, we evaluated intra-observer variability in semi-automated 3D cardiac volumetric analysis using CCT in patients with LVADs.
Methods:
Ten clinically indicated CCTs were studied retrospectively from 9 patients with LVADs. 3D chamber quantification included left and right ventricles end-systolic and end-diastolic volumes (ESV, EDV); and left and right atrial ESV. Derived measurements included cardiac output (CO), ejection fraction (EF), and stroke volume (SV). Automated volumetric analysis was performed, and manual corrections were added when necessary. Absolute and relative differences, Bland-Altman plots, and interclass correlation coefficients (ICCs) were used to assess intra-observer reproducibility for these measurements.
Results:
Intra-observer reproducibility was excellent for volumetric (ICC >0.99) and derived data (ICC >0.91). Comparing right vs left heart volumetric assessments, the former had a higher relative difference (atria 2.8% vs 1.6%, ESV 3.0% vs 1.9%, EDV 2.7% vs 1.3%), which also translated to a greater relative difference in right-side derived data (CO 11.1% vs. 8.8%, EF 10.5% vs. 9.9%, SV 10.9% vs. 9.0%). The mean difference in left ventricular ejection fraction was 0.4% (limits of agreement [LOA]: -2 and 3.2) and right ventricular ejection fraction was 1.2% (LOA: -4.7 and 7.1).
Conclusions:
Our results for semi-automated 3D volumetric analysis showed excellent reproducibility for both volumetric and derived data.
Summary:
Electrocardiography-gated cardiac computed tomography with semi-automated volumetric analysis has excellent reproducibility in patients with left ventricular assist device making it imaging modality of choice for functional assessment in this patient population, where cardiac magnetic resonance imaging is contraindicated and transthoracic echocardiography may be limited by poor acoustic windows.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Dimensional Analysis
Conversion Factors and Dimensional Analysis
The unit...
Dimensional Analysis
In fluid mechanics, dimensional...
Dimensional Analysis
Dimensional analysis allows us to analyze and compare physical quantities on a...
Dimensional Analysis
Imaging Studies III: Computed Tomography