Dose of CT protocols acquired in clinical routine using a dual-layer detector CT scanner: A preliminary report
F van Ommen1, H W A M de Jong1, J W Dankbaar2
1Department of Radiology and Nuclear Medicine, UMC Utrecht, P.O. Box 85500, 3508 GA, Utrecht, Netherlands; Image Sciences Institute, UMC Utrecht, P.O. Box 85500, 3508 GA, Utrecht, Netherlands.
Purpose:
To assess the radiation dose associated with always-on dual-energy acquisitions in clinical practice over a broad range of clinical protocols using a dual-layer detector CT (DLCT; IQon spectral CT, Philips Healthcare) as compared to an otherwise technically equivalent single-layer detector CT (SLCT; Brilliance iCT, Philips healthcare).
Materials And Methods:
Dose-length-product data for consecutive examinations over a six-month period acquired with DLCT were retrospectively collected and compared to consecutive examinations from an SLCT. Imaging protocols were optimized for diagnostic image quality for each system prior to data collection. Dose reports of CT protocols that were used at least 50 times on both systems were collected. After exclusion of statistical outliers, protocols were evaluated with regard to reported dose levels.
Results:
In total, 4536 dose reports for DLCT and 5783 reports for SLCT were collected. All DLCT examinations were acquired at 120 kVp, enabling dual-energy analysis. With SLCT, 79% of examinations were acquired at 120 kVp, and 21% at 100/80 kVp. Protocols for 15 indications were used more than 50 times on both scanners. For seven protocols there was no significant difference between the two scanners (p > 0.05), whereas seven protocols were acquired with higher dose levels on SLCT compared to the DLCT (p < 0.03). For one protocol, the DLCT dose was significantly higher (p < 0.005) compared to the SLCT.
Conclusion:
Dual-layer detector CT enables acquisition of dual-energy information over a broad range of clinical indications without increasing radiation dose when compared to a conventional single-layer detector CT.
More Related Videos
07:54Implantation and Monitoring by PET/CT of an Orthotopic Model of Human Pleural Mesothelioma in Athymic Mice
Published on: December 21, 2019
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...