Related Experiment Video
Updated: Jan 28, 2026

Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
Evaluation of the i-STAT analyzer for determination of ionized calcium concentrations in bovine blood
Oznur Yilmaz1, Tolga Karapinar1
1Department of Internal Medicine, Faculty of Veterinary Medicine, Firat University, Elazig, Turkey.
Background:
Information regarding the performance of the i-STAT, a point-of-care analyzer, to determine ionized calcium (iCa2+ ) concentration in bovine blood is limited.
Objective:
The primary goal of this study was to evaluate the agreement of the i-STAT against a traditional bench-top blood gas analyzer (Radiometer ABL800 FLEX) in the measurement of iCa2+ in the blood of cattle with various disorders.
Methods:
In this study, 121 cattle with various diseases were used. Venous blood samples were run on the i-STAT and the Radiometer ABL800 FLEX for determination of blood iCa2+ concentrations. Data were subjected to Passing-Bablok regression and Bland-Altman plots. The Radiometer ABL800 FLEX was the reference method and the i-STAT was the test method. The i-STAT performance was evaluated at two different cutoff values for iCa2+ concentrations (<1.00 and <1.18 mmol/L) using the receiver operating characteristic curve.
Results:
The Passing-Bablok regression equation that fit the iCa2+ concentration as measured with the i-STAT (Y) and the Radiometer ABL800 FLEX (X), was Y = -0.108276 [-0.15/-0.06] + 1.103448 [1.06/1.14] × X, with a residual deviation of 0.02. The deviation from the identity line was not significant (P = 0.64). The Bland-Altman difference plot revealed no bias between the i-STAT and the Radiometer ABL 800 FLEX. The i-STAT had a sensitivity of 100% (95% CI, 87.7-100.0) and 94.8% (95% CI, 87.2-98.6) at the iCa2+ concentration cutoff values of <1.00 and <1.18 mmol/L, respectively.
Conclusions:
iCa2+ concentrations in bovine blood measured with the i-STAT agreed with those of the Radiometer ABL800 FLEX. In particular, the i-STAT predicted hypocalcemia with a high sensitivity.
Related Concept Videos
Ionization Energy
Feedback Regulation of Calcium Concentration
Various transmembrane receptors, such as G protein-coupled receptors (GPCRs), elicit a response to extracellular signals by increasing cytosolic calcium. Activated GPCRs...
The JAK-STAT Signaling Pathway
Concentration Cells
Consider the following voltaic cell:
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Calculating Equilibrium Concentrations
A more...