A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Coronary vein defibrillator coil placement in patients with high defibrillation thresholds
Moisés Rodríguez-Mañero1,2,3,4, Bahij Kreidieh1, Sergio H Ibarra-Cortez1
1Methodist DeBakey Heart and Vascular Center and Methodist Hospital Research Institute The Methodist Hospital Houston Texas.
Background:
Elevated defibrillation threshold (DFT) occurs in 2%-6% of patients undergoing implantable cardioverter defibrillator (ICD) implantation. Adding a defibrillation coil in the coronary sinus (CS) or its branches can result in substantial reductions in the mean DFT. However, data regarding acute success and long-term stability remain lacking. We report our experience with this bailout strategy.
Methods:
Patients with elevated DFT at implantation (safety margin at implantation <10 J) and those with failed ICD shocks for ventricular arrhythmias (VA) referred for high DFT underwent placement of an additional defibrillation coil in the CS. DFT testing was performed at the completion of the implantation procedure. External potentially reversible factors were excluded. High-output devices were systematically used.
Results:
Four patients with high DFT at implantation and two with several failed shock attempts underwent placement of a defibrillation coil in the CS. Mean age was 41.8 (23-78). They presented a mean LVEF of 21% (15-30), QRS-complex duration of 109.8 milliseconds (87-168), body surface area of 1.96 m2 (1.45-2.58), and a mean R wave of 16.3 mV (8-27). Defibrillation coil implantation in the CS (final shocking configuration of right ventricle as anode and left ventricle (LV) plus can as cathode) was associated with successful DFT testing in all. Three patients had a concomitant LV lead for biventricular pacing. During a mean follow-up of 54.67 months (10-118), two patients experienced successful ICD shocks for VA (one of them also presented inappropriate shocks because of the fast conducting atrial fibrillation).
Conclusions:
Positioning of a defibrillation coil in the CS can result in a substantial reduction in mean DFT and associates with optimal long-term stability.
More Related Videos
05:52Myocardial Infarction by Percutaneous Embolization Coil Deployment in a Swine Model
Published on: November 4, 2021
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
Related Concept Videos
Coronary Circulation
Coronary circulation begins at the base of the aorta, where two main arteries arise—the left and right coronary arteries. These arteries encircle the heart in the coronary sulcus and supply the...
Veins
Structure of Veins:
The structure of veins is specifically designed to assist in the low-pressure transportation of...
Veins of Thorax
The azygos vein, positioned just right of the midline and anterior to the vertebral column, begins at the junction of the right ascending lumbar and subcostal veins, terminating in the superior vena cava. This vein drains blood from the right side of the thoracic wall, thoracic viscera, and posterior abdominal wall.
The...
Veins of Lower Limbs
Formed by the union of the medial and lateral plantar veins, the posterior tibial vein, rising through the calf muscle, assimilates the fibular vein. The anterior tibial vein, a superior extension of the foot's dorsalis pedis vein, merges with the posterior tibial vein at the...
Coronary Artery Disease I: Introduction
Coronary Artery Disease II: Pathophysiology