The efficacy of tin-filtration for computed tomography in diagnosing urolithiasis
Sarah Li Hui Tan1, Mohamed Badawy2, Michal Schneider1
1Department of Medical Imaging and Radiation Sciences, Monash University, Australia.
Introduction:
The purpose of this study was to evaluate the radiation dose and image quality of computed tomography urograms (CTU) using tin-filtration compared to conventional CTU (without tin-filtration) examinations in patients with suspected urolithiasis.
Methods:
Group 1 consisted of 100 patients who were examined using the tin-filtered CTU protocols (Sn100kVp or Sn150kVp); Group 2 consisted of 100 patients who were examined using the same protocols but without tin-filtration (GE-NI41 or GE-NI43). The scanning protocol was based on the patients' body weight (<80 kg and ≥80 kg). The effective doses of all scans were compared between the two groups. Subjective image quality was evaluated by two blinded radiologists. The objective image quality was assessed for noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and figure-of-merit (FOM) using the CTU scans acquired from both the tin-filtered and non-tin-filtered protocols.
Results:
Tin-filtration resulted in the reduction of effective radiation dose ranging between 72% to 88% for the ≥80 kg and <80 kg patient groups respectively. For both groups, tin-filtration resulted in no significant differences in SNR and a significant increase in FOM. For the <80 kg group, tin-filtration resulted in significantly noisier images but with no significant difference in CNR. For the ≥80 kg group, tin-filtration resulted in significantly higher CNR. There was no significant difference in subjective image quality when assessed by the radiologists in terms of diagnostic confidence for urolithiasis.
Conclusion:
Tin-filtration significantly reduces patient dose while maintaining diagnostic image quality of CTUs for patients with suspected urolithiasis.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Filtration and Urine Formation
Glomerular Filtration: Net Filtration Pressure
GBHP, with an average value of 55 mmHg, promotes filtration by pushing water and solutes through the filtration membrane. This is balanced by two opposing forces: CHP, a "back pressure" exerted against the filtration membrane by fluid already in the capsular space and renal...
Filtration
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and...