Published on: February 20, 2021
A novel risk analysis of clinical reference dosimetry based on failure modes and effects analysis
Yusuke Ochi1, Akito Saito2, Daisuke Kawahara1
1Radiation Therapy Section, Department of Clinical Support, Hiroshima University Hospital, Hiroshima 734-8551, Japan; Department of Radiation Oncology, Graduate School of Biomedical & Health Sciences, Hiroshima University, Hiroshima 734-8551, Japan.
Purpose:
The output of a linear accelerator (linac) is one of the most important quality assurance (QA) factors in radiotherapy. However, there is no quantitative rationale for frequency and tolerance. The purpose of this study is to develop a novel risk analysis of clinical reference dosimetry based on failure modes and effects analysis (FMEA).
Methods:
Clinical reference dosimetry data and the daily output data of two linacs (Clinac iX and Clinac 6EX) at Hiroshima University Hospital were analyzed. The analysis involved the number of patients per year for five types of fractionations. Risk priority number (RPN) is defined as the product of occurrence (O), severity (S), and detectability (D) in standard FMEA. In addition, we introduced "severity due to output drifting" (mean output change per day) (S') and the number of patients per year for five types of fractionations (W). We calculated the RPN = O × S × D × S' × W and quantitatively evaluated the risk for clinical reference dosimetry.
Results:
Fewer fractions and less output calibration frequency resulted in higher RPN. Since clinical reference dosimetry data has a drift effect, which is missing in human processes, it was essential to use S' in addition to standard FMEA. Moreover, the parameter W was important in evaluating interinstitutional QA for clinical reference dosimetry. The relative risk of Clinac 6EX to Clinac iX was different approximately by twofold.
Conclusions:
We developed a novel index that can quantitatively evaluate risk for clinical reference dosimetry of each facility and machines in common on the basis of FMEA.
More Related Videos
Related Concept Videos
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Heart Failure III: Clinical Manifestations
Statistical Software for Data Analysis and Clinical Trials
Pedigree Analysis
Qualitative Analysis
For instance, group IV...
What is a Mode?
There can be more than one mode in a data set if multiple values have the same highest frequency. For instance, suppose that the Statistics exam scores of 20 students are: 50; 53; 59; 59; 63; 63; 72; 72; 72; 72; 72; 76; 78; 81; 83; 84; 84; 84; 90; 93. Here, the mode is 72, as it occurs most frequently, five times.
A data set with two modes is called bimodal. For example,...