Related Experiment Video
Updated: Jan 28, 2026

Author Spotlight: Development of a Minimally Invasive Large-Animal Model for Reliable and Reproducible Cardiovascular Research
Published on: October 20, 2023
Minimally invasive percutaneous epicardial placement of a prototype miniature pacemaker with a leadlet under direct
Rohan N Kumthekar1, Justin D Opfermann2, Paige Mass2
1Division of Cardiology, Children's National Health System, Washington, District of Columbia.
Background:
Pacemaker implantation in infants is limited to epicardial lead placement and an abdominal generator pocket. We propose a minimally invasive solution using a prototype miniature pacemaker with a steroid-eluting leadlet that can affix against the epicardium under thoracoscopy.
Objective:
The purpose of this study was to evaluate the safety and feasibility of acute implantation of a prototype miniature pacemaker in an infant porcine model.
Methods:
A self-anchoring 2-channel access port was inserted into a 1-cm incision left of the subxiphoid space. A rigid thoracoscope with variable viewing angle was inserted through the main channel to visualize the heart under insufflation. An 18-G needle through the second channel accessed the pericardial space, which was secured with a 7-F sheath. The leadlet was affixed against the epicardium using a distal helical side-biting electrode. The sheath, thoracoscope, and port were removed, and the pacemaker was tucked into the incision. Ventricular sensing, lead impedances, and capture thresholds were measured.
Results:
Twelve piglets (weight 4.8 ± 1.9 kg) had successful device implantation. The median time from incision to leadlet fixation was 21 minutes (interquartile range [IQR] 18-31 minutes). The median lead impedance was 510 Ω (IQR 495-620 Ω). The median R-wave amplitude was 5.7 mV (IQR 4.2-7.0 mV). The median capture threshold was 1.63 V (IQR 1.32-2.97 V) at 0.4 ms pulse width and 1.50 V (IQR 1.16-2.38 V) at 1.0 ms pulse width. There were no complications.
Conclusion:
Minimally invasive epicardial placement of a prototype miniature pacemaker under thoracoscopy was safe and avoided open chest surgery and creation of an abdominal generator pocket.
Related Concept Videos
Concepts and Prototypes
The brain organizes this information using concepts, which are mental categories grouping linguistic data,...
Drug Dosing: Infants and Children
Directing Effect of Substituents: meta-Directing Groups
Typical Model Studies
Directing Effect of Substituents: ortho–para-Directing Groups
Mouse Models of Cancer Study
The development of transgenic, knockout, and knock-in mice has led to an exponential increase in their use as model organisms in research,...