Protocol for Developing a Femur Osteotomy Model in Wistar Albino Rats
Published on: August 31, 2022
Osteotomy speed, heat development, and bone structure influence by various piezoelectric systems-an in vitro study
Nicolas Noetzel1, Tim Fienitz2, Matthias Kreppel3
1Department for Oral and Cranio-Maxillo and Facial Plastic Surgery, University of Cologne, Kerpener Straße 62, 50937, Cologne, Germany. nicolas.noetzel@gmx.de.
Objectives:
The aim of this in vitro study was to evaluate osteotomy speed, heat development, and bone structure influence from osteotomies performed by various piezoelectric devices and insert tips. These devices and tips were compared among each other with regard to conventional rotatory and oscillating systems with special focus on the insert tip design and thickness.
Material And Methods:
The osteotomies were conducted on porcine ribs utilizing 12 different insert tips (straight and angulated) and three conventional systems. After time and temperature measurements, histological analysis was carried out. Light microscopy was used to evaluate the roughness of the osteotomic surface and to search for indications of thermal bone necrosis. A special software analyzing tool was employed to determine cutting width (mm) and debris (%).
Results:
All piezoelectric tips created smooth cuts. Cutting widths in general were wider than the actual insert tip size with a tendency for narrow straight insert tips producing relatively wide osteotomies, whereas narrow angulated inserts produced relatively small osteotomies. None of the samples demonstrated distinct indication of necrosis. Overall, there was only a small amount of debris in all osteotomy gaps. Conventional rotatory saws were faster and created less heat compared to all tested piezoelectric systems. Straight tips proved faster osteotomy speed than angulated tips. Thin insert tips indicated to have a positive correlation to osteotomy time and performed faster than conventional microsaw. The average temperature rise was lower when using conventional systems, but critical exceeding temperatures were only observed in short-time exceptional cases. In general, temperature rise was less when using angulated inserts.
Conclusion:
All tested tips are appropriate for bone surgery. Only small differences were found among the piezoelectric insert tips. Although conventional rotatory systems in general performed faster osteotomies, special designed and thin piezoelectric insert tips seem to have a positive influence on osteotomy speed. Ultimately, none of the tested devices or inserts combined all best features of speed, heat development, bone structure influence, and safety.
Clinical Relevance:
Narrow and straight piezoelectric insert tips demonstrated reduced osteotomy times. Nevertheless, a combination of conventional and piezoelectric systems in clinical practice might be the best way to work time-efficient, patient-oriented, and safe. The choice of instrument should be based on clinical experience of the user and should be evaluated individually depending on the case.
More Related Videos
10:09Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
07:41An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
Related Concept Videos
Bone Structure
Fruit Development, Structure, and Function
Seed Structure and Early Development of the Sporophyte
Specific Heat
For example, increasing the temperature of one gram of water by 1°C requires one calorie of heat energy and can be written as 1 cal/g-°C, or...
Responses to Heat and Cold Stress
Quantifying Heat