Comparative Cardiac Phantom Study Using Tc-99m/I-123 and Tl-201/I-123 Tracers with Cadmium-Zinc-Telluride
Takanaga Niimi1, Mamoru Nanasato2, Mitsuo Sugimoto1
11Department of Radiological Technology, Nagoya Daini Red Cross Hospital, 2-9 Myouken-cho, Showa-ku, Nagoya, 466-8650 Japan.
Objective:
A recently introduced single-photon emission computed tomography (SPECT), based on cadmium-zinc-telluride (CZT) detectors (D-SPECT), supports high energy resolution for cardiac imaging. Importantly, the high energy resolution may allow simultaneous dual-isotope (SDI) imaging (e.g., using Tc-99m and I-123). We quantitatively evaluated Tc-99m/I-123 SDI imaging by D-SPECT in comparison with conventional T1-201/I-123.
Materials And Methods:
Energy resolution was measured as a percentage of the full width at half maximum (FWHM) for Tc-99m, I-123, and Tl-201. The impact of cross-talk and reconstructed image contrast were quantified by measuring the contrast-to-noise ratio (CNR), and the transmural defect contrast in the left ventricle wall (C TD) induced by a difference in energy, for combinations of Tc-99m/I-123 or Tl-201/I-123, using an RH-2 cardiac phantom. Corresponding measurement was also carried out in Anger SPECT (A-SPECT).
Results:
The energy resolution of the D-SPECT system was 5.4%/5.1% for Tc-99m/I-123 and 5.4%/5.3% for Tl-201/I-123, which was approximately two times higher than the A-SPECT. No notable difference was confirmed in the CNRs of the two systems, but T1-201/I-123 showed overall higher value than Tc-99m/I-123. Compared to A-SPECT, C TD of D-SPECT significantly increased with both Tc-99m/I-123 and T1-201/I-123 (p < 0.05). In DSPECT, the combination of Tc-99m/I-123 had a slightly better C TD than T1-201/I-123. In addition, C TD of Tc-99m/I-123 was improved with scatter correction at both nuclides (p < 0.05), but in Tl-201/I-123, no significant improvement was confirmed in I-123 (p > 0.05).
Conclusion:
D-SPECT was considered to be capable of performing high-quality SDI imaging using Tc-99m/I-123.
More Related Videos
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
06:52Positron Emission Tomography Using 64-Copper as a Tracer for the Study of Copper-Related Disorders
Published on: April 28, 2023
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Emission Spectra
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...