Polymethylmethacrylate Collagen Gel-Injectable Dermal Filler for Full Face Atrophic Acne Scar Correction
John H Joseph1, Ava Shamban2, Laura Eaton3
1Clinic Testing of Beverly Hills, Beverly Hills, California.
Background/Objectives:
This trial evaluated the effectiveness and safety of Bellafill for full-face acne scar treatment.
Patients And Methods:
In this open-label, nonrandomized, multicenter pilot study investigating the use of polymethylmethacrylate for full-face atrophic acne scar correction, 42 adult subjects with a mean age of 43 years were treated and assessed for safety and effectiveness at Months 4 and 7. There were no hypersensitivity reactions to pretreatment skin testing or during scar treatments.
Results:
At 4 and 7 months after initial treatment, 92% and 95% of subjects, respectively, were responders with ≥1-point improvement on the 5-point Acne Scar Assessment Scale. Subjects reported very high levels of improvement on the Global Aesthetic Improvement Scale (GAIS), with 95% of subjects reporting "improved or better" at 4 months and 90% at 7 months. The outcome of the physician GAIS was also high with 92% of patients classified as "improved or better" at 4 months and 97% at 7 months. There were only 2 device-related adverse events, both mild events related to Bellafill skin test (bruising, ecchymosis). There were no serious adverse events in response to the treatment product in this short-term follow-up study.
Conclusion:
Polymethylmethacrylate is effective for treating full-face acne scarring. Clinicaltrials.gov #NCT02642627.
Related Concept Videos
Additives and Fillers in Concrete
The...
Distance Corrections
Power Factor Correction
Fibril-associated Collagen
For example, the type II collagen fibrils in cartilage have covalently bound type IX fibril-associated collagens at regular intervals. Other types of fibril-associated collagens are...
NMR Spectrometers: Resolution and Error Correction
Two-dimensional Gel Electrophoresis
The first dimension separation uses the isoelectric focusing or IEF technique performed on immobilized pH gradient (IPG) strips that separate proteins according to their isoelectric points.
Biological samples, such...