Related Experiment Video
Updated: Jan 28, 2026

Protein Isolation from the Developing Embryonic Mouse Heart Valve Region
Published on: September 23, 2014
Assessing the Thrombogenic Potential of Heart Valve Prostheses: An Approach for a Standardized In-Vitro Method
Torsten Linde1, Johanna Clauser2, Bart Meuris3
1Department of Cardiovascular Engineering, Helmholtz Institute of Applied Medical Engineering, RWTH Aachen University, Aachen, Germany. tlinde@netaachen.de.
Purpose:
Thrombogenic complications are still a main issue in the general performance of cardiovascular implants, especially heart valves. To date, the thrombogenic potential of those prostheses is pre-clinically assessed in time consuming animal studies with questionable evidence.
Methods:
In this study, we present a new in-vitro method to assess and compare deficiencies of heart valve substitutes concerning their thrombogenic performance and locate initial clot formation under physiological conditions using porcine blood. Therefore, an athrombogenic pulse duplicator (THIA3) was developed that simulates the anatomic and hemodynamic conditions in the vicinity of the aortic valve. Validation of this tester was carried out with regard to hemodynamics, reproducibility and using positive and negative control valves and by comparison of clot locations with literature data from chronic animal trials with sheep using a St. Jude bileaflet valve.
Results:
Validation of the tester showed quasi-physiological hemodynamics and reproducible clot formation. Identical clot formations were found comparing findings in vitro with chronic animal trials.
Conclusion:
The THIA 3 has proven its suitability for valid, reproducible evaluation of thrombogenic potential of heart valves in a short period of time.
Related Concept Videos
Standard Electrode Potentials
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Valve Prolapse II: Assessment and Management
Anatomy of the Heart
Empirical Method to Interpret Standard Deviation
This rule is used widely in statistics to calculate the proportion of data values...
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential...