Related Experiment Video
Updated: Jan 28, 2026

Determining The Electromyographic Fatigue Threshold Following a Single Visit Exercise Test
Published on: July 27, 2015
Influence of proper or reciprocating optimum torque reverse kinematics on cyclic fatigue of four single files
Eugenio Pedullà1, Emanuele Ambu2, Fabio Rovai3
1Department of General Surgery and Surgical-Medical Specialties, University of Catania, Catania, Italy.
Aim:
The aim of the present study was to evaluate the resistance to cyclic fatigue of heat-treated single files used in proper or in reciprocating optimum torque reverse motions.
Methods:
Ninety-six HyFlex electrical discharge machining (EDM), Reciproc R25 (Rec), WaveOne Primary (WO), and WaveOne Gold (WOG) Primary were used. Resistance to cyclic fatigue was determined by recording time to fracture in a stainless steel artificial canal. Differential scanning calorimetric analysis was performed. Data were analyzed by two-way analysis of variance and Bonferroni post-hoc tests at 0.05.
Results:
In both movements, HyFlex EDM showed higher time to fracture than other files (P < 0.0001). Rec showed higher time to fracture than WO (P < 0.0001), while WOG showed higher cyclic fatigue resistance than Rec and WO with both movements (P < 0.005).
Conclusions:
Reciprocating optimum torque reverse motion improved significantly cyclic fatigue resistance of gold and controlled memory wire single files.