Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Diffusion01:12

Diffusion

217.8K
Diffusion is the passive movement of substances down their concentration gradients—requiring no expenditure of cellular energy. Substances, such as molecules or ions, diffuse from an area of high concentration to an area of low concentration in the cytosol or across membranes. Eventually, the concentration will even out, with the substance moving randomly but causing no net change in concentration. Such a state is called dynamic equilibrium, which is essential for maintaining overall...
217.8K
Diffusion01:21

Diffusion

6.4K
Diffusion is a type of passive transport. In passive transport, a substance tends to move from an area of high concentration to an area of low concentration until the concentration is equal across the space. For example, take the diffusion of substances through the air. When someone opens a perfume bottle in a room filled with people, the perfume is at its highest concentration in the bottle and is at its lowest at the edges of the room. The perfume vapor will diffuse, or spread away, from the...
6.4K
Microsoft Excel: Finding Central Tendency, Skew, and Kurtosis01:24

Microsoft Excel: Finding Central Tendency, Skew, and Kurtosis

663
Central tendency refers to the central point or typical value of a dataset. It summarizes the data set with a single value that represents the center of its distribution. The three main measures of central tendency are:
Mean: The arithmetic average of all data points. It is calculated by adding all the values together and dividing by the number of values. The mean is sensitive to extreme values (outliers).
Median: The middle value when the data points are arranged in ascending or descending...
663
Facilitated Diffusion01:16

Facilitated Diffusion

1.3K
The plasma membrane, a critical structure in cellular biology, houses an array of transporters, or carrier proteins, interspersed within its lipid bilayer. These proteins play a crucial role in solute transport through facilitated diffusion, a form of passive diffusion that uses transporters to move the molecules across the membrane.
In this process, substrates such as organic compounds and ions interact with a transporter on one side, triggering conformational changes in proteins that enable...
1.3K
Imaging Studies VII: Vascular Imaging01:19

Imaging Studies VII: Vascular Imaging

360
DefinitionRenal angiography, also known as renal arteriography, is an imaging technique used to obtain a comprehensive view of blood flow and the vascular structure of blood vessels in the kidneys and surrounding areas.PurposeRenal angiography detects blood vessel abnormalities in the kidneys, such as aneurysms, stenosis, thrombosis, vascular tumors, and renal artery stenosis. It evaluates kidney function and guides interventional treatments like angioplasty or stent placement.Pre-Procedure...
360
Imaging Studies II: Ultrasonography01:24

Imaging Studies II: Ultrasonography

419
IntroductionUltrasonography, or renal ultrasound, is a noninvasive medical imaging technique that uses high-frequency sound waves to visualize the kidneys, ureters, bladder, and surrounding tissues.Indications for Urinary System UltrasonographyUrinary system ultrasonography is indicated in various clinical scenarios, such as:Kidney Stones (Urolithiasis): To detect and monitor the size and presence of kidney or urinary tract stones.Hydronephrosis: To assess the dilation of the renal pelvis and...
419

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

CAR T-Cell Therapy for Cancer: Updates and Challenges for Response Assessment.

Journal of nuclear medicine : official publication, Society of Nuclear Medicine·2026
Same author

AI-MIRACLE: Artificial Intelligence and MultIpaRAmetric MRI Predict CLinical OutcomEs to Neoadjuvant Immunotherapy in Patients with Muscle-invasive Bladder Cancer Undergoing Radical Cystectomy.

European urology oncology·2026
Same author

AI-BLADE toolbox: AI-powered BLADdEr multiparametric MRI analysis for clinical application.

BJR artificial intelligence·2026
Same author

Optimizing Colorectal Liver Metastasis Ablation through Metabolic Imaging, Margin Assessment, and Biopsy (The OPTABLATE Prospective Trial).

Radiology·2026
Same author

Towards Clinical Translation of Intravoxel Incoherent Motion MRI: Acquisition and Analysis Consensus Recommendations.

Journal of magnetic resonance imaging : JMRI·2026
Same author

Reply to the Letter to the Editor: A deep learning framework to stratify Nottingham histologic grade 2 breast tumors based on dynamic contrast-enhanced MRI.

European radiology·2026
Same journal

Age-Related Concentric Remodeling and Sex-Dependent Dimensional Variation in Left Ventricular Geometry: A Cardiac Magnetic Resonance Study.

Tomography (Ann Arbor, Mich.)·2026
Same journal

Opportunistic Screening for Low Bone Density Using Automated Vertebral Trabecular CT Attenuation from Low-Dose CT Acquired During FDG PET/CT: A Single-Center Retrospective Study.

Tomography (Ann Arbor, Mich.)·2026
Same journal

Machine Learning-Based Classification of BI-RADS 4 and BI-RADS 5 Microcalcifications in Mammography Combined with DCE-MRI for Malignant-Benign Discrimination.

Tomography (Ann Arbor, Mich.)·2026
Same journal

Image Quality Assessment of Diffusion-Weighted Imaging (DWI) and Its Impact on Apparent Diffusion Coefficient (ADC) as a Quantitative Imaging Biomarker for Predicting Response to Neoadjuvant Chemotherapy in High-Risk Early Breast Cancer.

Tomography (Ann Arbor, Mich.)·2026
Same journal

Relationship Between Cervical Central Canal and Neural Foraminal Dimensions in a Normative Population.

Tomography (Ann Arbor, Mich.)·2026
Same journal

AI-Based Scientific Manuscript Peer Review: Is It Ready for Adoption?

Tomography (Ann Arbor, Mich.)·2026
See all related articles

Related Experiment Video

Updated: Jan 28, 2026

Advanced Diffusion Imaging in The Hippocampus of Rats with Mild Traumatic Brain Injury
10:33

Advanced Diffusion Imaging in The Hippocampus of Rats with Mild Traumatic Brain Injury

Published on: August 14, 2019

9.0K

Multicenter Repeatability Study of a Novel Quantitative Diffusion Kurtosis Imaging Phantom.

Dariya I Malyarenko1, Scott D Swanson1, Amaresha S Konar2

  • 1Department of Radiology, University of Michigan Medical School, Ann Arbor, MI.

Tomography (Ann Arbor, Mich.)
|March 12, 2019
PubMed
Summary
This summary is machine-generated.

This study evaluates a new physical model designed to standardize measurements in advanced brain imaging. By using liquid crystal materials, researchers created a stable tool to ensure that different hospitals produce consistent results when measuring complex water movement in tissues. The findings demonstrate that this new device provides reliable data over time, supporting its use in large-scale clinical trials to improve diagnostic accuracy.

Keywords:
diffusion kurtosismicro-scale lamellar vesiclesrepeatabilitytemporal stabilitytunable parametersMagnetic Resonance ImagingBiomarker StandardizationMedical PhysicsImaging Precision

Frequently Asked Questions

More Related Videos

Agarose-based Tissue Mimicking Optical Phantoms for Diffuse Reflectance Spectroscopy
09:25

Agarose-based Tissue Mimicking Optical Phantoms for Diffuse Reflectance Spectroscopy

Published on: August 22, 2018

13.2K
Author Spotlight: A Stable Phantom Material for Optical and Acoustic Imaging
04:54

Author Spotlight: A Stable Phantom Material for Optical and Acoustic Imaging

Published on: June 16, 2023

3.8K

Related Experiment Videos

Last Updated: Jan 28, 2026

Advanced Diffusion Imaging in The Hippocampus of Rats with Mild Traumatic Brain Injury
10:33

Advanced Diffusion Imaging in The Hippocampus of Rats with Mild Traumatic Brain Injury

Published on: August 14, 2019

9.0K
Agarose-based Tissue Mimicking Optical Phantoms for Diffuse Reflectance Spectroscopy
09:25

Agarose-based Tissue Mimicking Optical Phantoms for Diffuse Reflectance Spectroscopy

Published on: August 22, 2018

13.2K
Author Spotlight: A Stable Phantom Material for Optical and Acoustic Imaging
04:54

Author Spotlight: A Stable Phantom Material for Optical and Acoustic Imaging

Published on: June 16, 2023

3.8K

Area of Science:

  • Medical physics and quantitative Diffusion Kurtosis Imaging phantom validation
  • Biomedical engineering within diagnostic imaging research

Background:

Standardizing advanced magnetic resonance imaging metrics across different clinical sites remains a significant challenge for modern medicine. No prior work had resolved the need for reliable physical models to calibrate complex diffusion measurements. That uncertainty drove the development of specialized tools to ensure consistency in multicenter trials. Prior research has shown that current methods often struggle with inter-scanner variability during longitudinal assessments. This gap motivated the creation of a stable, isotropic physical reference for validating specific imaging parameters. Researchers have long sought reliable ways to distinguish between true biological signals and technical noise in these scans. Establishing such benchmarks is vital for the widespread adoption of quantitative biomarkers in clinical practice. This study addresses these limitations by introducing a novel design based on liquid crystal mesophases.

Purpose Of The Study:

The aim of this work is to evaluate the precision and long-term stability of a newly developed physical reference for advanced imaging. Researchers sought to address the lack of standardized tools for calibrating complex diffusion measurements in multicenter trials. They specifically focused on creating an isotropic phantom that mimics biological tissue characteristics. This initiative was motivated by the need to ensure that quantitative biomarkers remain consistent across different hospital scanners. The study investigates whether liquid crystal mesophases can provide a reliable foundation for such calibration devices. By testing the prototype over several months, the authors intended to establish its utility for clinical research. The project also explores how to effectively characterize noise and model bias within these imaging parameters. This effort provides a necessary framework for improving the reproducibility of quantitative diagnostic data.

Main Methods:

The review approach involved a longitudinal assessment of a newly engineered physical reference tool. Investigators utilized four distinct chemical families to construct the isotropic prototype. They conducted ten repeated measurements across four separate magnetic resonance scanners. These sessions occurred at three distinct clinical facilities over a half-year duration. The team performed correlation analysis to evaluate inter-scan variability and temporal stability. They also employed negative control materials to isolate model-induced bias from actual kurtosis signals. Statistical evaluation focused on calculating 95% confidence intervals for the primary imaging parameters. This rigorous testing framework ensured that the phantom performance remained consistent under varying operational conditions.

Main Results:

Key findings from the literature reveal that the prototype successfully produced physiologically relevant values for both diffusion and kurtosis parameters. The apparent diffusion coefficient ranged from 0.4 to 1.1 (×10⁻³ mm²/s), while kurtosis values spanned 0.8 to 1.7. The measured kurtosis consistently exceeded the maximum fit model bias of 0.1 observed in the negative control samples. Statistical analysis yielded a 95% confidence interval for diffusion between 0.013 and 0.022 (×10⁻³ mm²/s). The corresponding interval for kurtosis precision was determined to be between 0.009 and 0.076. These results confirm that the device maintains sufficient stability for characterizing thermal and temporal fluctuations. The data demonstrate that the chemical design effectively minimizes variability across different imaging centers. This performance indicates that the phantom is suitable for validating quantitative biomarkers in clinical settings.

Conclusions:

The authors demonstrate that their liquid crystal design offers a robust solution for standardizing diffusion measurements across multiple sites. This synthesis suggests that the prototype maintains sufficient stability for use in long-term clinical monitoring. The findings imply that the observed precision levels are adequate for distinguishing true kurtosis signals from model-induced noise. The researchers propose that refined preparation techniques will further improve the reproducibility of these physical models. Future efforts should focus on integrating precise temperature monitoring to enhance the reliability of the measurements. The study confirms that the vesicular mesophase approach provides a viable pathway for creating consistent reference standards. These results indicate that the current phantom design effectively supports the requirements of multicenter imaging research. The evidence supports the utility of these tools in validating quantitative biomarkers for clinical application.

The researchers propose that the device utilizes liquid crystal materials in vesicular and lamellar mesophases. This specific chemical composition allows the phantom to mimic physiologically relevant diffusion and kurtosis values, providing a stable reference for calibrating scanners across different clinical locations.

The team incorporated negative control materials that exhibit monoexponential diffusion. These components are necessary to independently quantify the impact of background noise and potential bias introduced by the mathematical models used to calculate kurtosis parameters.

The authors state that the phantom must be evaluated over a six-month period across multiple scanners. This duration is required to characterize the thermal and temporal stability of the device, ensuring that the measurements remain consistent despite environmental fluctuations.

The study utilizes test-retest data collected from ten separate imaging sessions. This approach allows the researchers to derive 95% confidence intervals for diffusion and kurtosis parameters, which are essential for determining the precision of the phantom.

The prototype demonstrated apparent diffusion values between 0.4 and 1.1 (×10⁻³ mm²/s) and kurtosis values between 0.8 and 1.7. These ranges are physiologically relevant, making the phantom suitable for validating clinical imaging protocols.

The researchers propose that future studies should implement improved temperature monitoring procedures. They suggest that these enhancements will increase the overall reproducibility of the phantom, making it more effective for large-scale multicenter trials.