Variations in intraocular lens injector dimensions and corneal incision architecture after cataract surgery
Alejandro Arboleda1, Esdras Arrieta1, Mariela C Aguilar1
1Ophthalmic Biophysics Center, Bascom Palmer Eye Institute, Department of Ophthalmology, University of Miami Miller School of Medicine, Jacksonville, Florida, USA.
Purpose:
To evaluate the differences in intraocular lens (IOL) injectors and to assess the effect of IOL insertion on injector tips and eyes after cataract surgery in a rabbit model.
Setting:
Ophthalmic Biophysics Center, Bascom Palmer Eye Institute, Department of Ophthalmology, University of Miami Miller School of Medicine, Miami, Florida, USA.
Design:
Experimental study.
Methods:
A modified optical comparator was used to measure the tips of 13 IOL injector models to determine the perimeter, tip angle, and cone angle of each. Injectors were analyzed before and after IOL insertion. Surgery was performed on rabbits with 71 IOL injectors of 13 models, and custom gauges were used to determine the incision size before and after surgery.
Results:
The injector dimensions varied by model; tip diameter, tip angle, and cone angle ranged from 1.44 to 2.12 mm, 29.7 to 66.5 degrees and 0.6 to 10.8 degrees, respectively. The incision size through which surgery was successfully performed also varied by injector model; the initial incision sizes ranged from 2.0 to 2.63 mm. For all injectors, there was wound enlargement after IOL insertion that ranged from a 0.1 to 0.65 mm increase in incision length.
Conclusions:
The dimensions and injection systems varied with each IOL injector. All injectors led to postoperative wound stretch after IOL insertion, with no final incision measuring less than 2.0 mm. These findings suggest that the clear cornea incision should have a width corresponding to the injector diameter.
Related Concept Videos
What is Variation?
The range, standard deviation, standard error, and variance are the different measures of variation.
Range: The range is the difference between its maximum and...
Conservative Site-specific Recombination and Phase Variation
The recognition sites for Cre recombinase called LoxP...
Variation
When independent and dependent variables are plotted on a scatter plot, the slope of a line is a value that describes the rate of change between the two...
Support Reactions in Three Dimensions
Ball and Socket Joint is one of the supports allowing free rotation about any axis. This freedom of rotation is...
Relative Velocity in One Dimension
Variation of Atmospheric Pressure
Assuming the air temperature is constant at a given altitude and that the ideal gas law of thermodynamics describes the atmosphere to a good approximation, one can find the variation of atmospheric pressure with height.
Let p(y) be the atmospheric pressure at...