Masimo Perfusion Index Versus Doppler for Tourniquet Effectiveness Monitoring
Background:
In addition to a plethysmograph, Masimo pulse oximeters display a Perfusion Index (PI) value. This study investigated the possible usefulness of PI for monitoring limb tourniquet arterial occlusion.
Methods:
Tactical Ratcheting Medical Tourniquets were applied to the thighs of 15 subjects. Tightening ended at one ratchet-tooth advance beyond Doppler- indicated occlusion. The times and pressures of Doppler and PI signal absences and returns were recorded.
Results:
Intermittent PI signal error occurred in 149 of 450 runs (PI, 33% versus Doppler, 0%; p < .0001). PI signal loss lagged Doppler-indicated occlusion by 19 ± 15 seconds (mean ± standard deviation, p < .0001). PI Signal Return lagged tourniquet release by 13 ± 7 seconds (Doppler Signal Return took 1 ± 1 seconds following tourniquet release; p < .0001). PI failed to detect early Doppler audible pulse return in 30 of 39 occurrences.
Conclusion:
The PI available on Masimo pulse oximeters is not appropriate for monitoring limb tourniquet effectiveness.
Related Concept Videos
Doppler Effect - I
Doppler Effect - II
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Buffer Effectiveness
The buffer capacity is the amount of acid or base that can be added to a given volume...
Framing Effects
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...